7.2. Replacing the fuse
WARNING!
To avoid electric shock, disconnect all the test probes before
removing the fuse. Replace only with the same type of fuse.
Service should be performed only by qualified personnel.
CAUTION!
For continued protection against fire or other hazard, replace only
with fuse of the specified voltage and current ratings.
Follow these steps to replace the fuse:
1. Turn ON/OFF button to turn the meter off and disconnect the
test
probes.
2. Remove the back cover by unscrewing the 6 screws and
pulling off the back cover.
3. Remove the blown fuse.
4. Install the new fuse in the fuse compartment.
FF 500mA/600V; 5 x 20mm
F 10A/600V; 6,3 x 32mm
5. Replace the battery compartment and secure it with the
screws.
WARNING!
Do not operate your meter until the back cover is in place and fully
closed.