- 31 -
6.5.4 Pulse width trigger
Pulse trigger occurs according to the width of pulse. The abnormal signals can be detected
through setting up the pulse width condition.
In Pulse Width Trigger mode, the trigger setting information is displayed on bottom right of the
screen, for example
,
indicates that trigger type is pulse width, trigger source
is CH1, coupling is DC, polarity is positive, and trigger level is 0.00mV.
Pulse trigger menü
:
Menu
Settings
Description
Trigger mode
Pulse
Set vertical channel trigger type as pulse trigger.
Source
CH1
CH2
CH3
CH4
Select CH1 as the trigger source
Select CH2 as the trigger source
Select CH3 as the trigger source
Select CH4 as the trigger source
Coupling
AC
DC
Not allow DC portion to pass.
Allow all portion pass.
When
Choose the polarity
Select pulse width condition and click
or
to set
time, click
to move cursor to choose which digit to
be set.
Mode
Holdoff
Auto
Normal
Single
Holdoff
Sensitivity
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, click
or
to set time interval before
another trigger occur, click
to move cursor to
choose which digit to be set
Set the trigger sensitivity
6.5.5 Slope trigger
The slope mode allows the oscilloscope to trigger on the rising / falling edge of a signal within a
specified period of time. In the slope trigger mode, the setting information is shown on the lower
right edge of the screen, e.g:
indicates that the slope trigger has been selected
on CH1, slope increasing and the difference between the up-level and low-level threshold value is
0.00mV.