Safely Isolated Condition
Definition: The unit is in a Safely Isolated Condition when it is disconnected from its
application, fully de-pressurised and isolated from the Electrical Supply. Directions for
isolating the Generator are shown below.
Failure to place the Generator in a safely isolated condition when
instructed to do so may lead to personal injury or injury to others and
even death.
Isolating the Generator:
The generator may be restored to a depressurised state by pressing and holding the
front button for at least three seconds. This will cause the generator to cease hydrogen
production, illuminate the water tank blue, and exhaust any stored hydrogen through
the exhaust port at the rear of the generator.
At this point you may switch the generator off at its rear power switch.
Safe State After Repair
To guarantee the generator is in a safe state after a repair, please ensure the following
conditions are met;
• The electricity cable has been correctly refitted.
The mains cable should be plugged into the IEC power inlet located on the rear
panel of the generator. This unit is classified as SAFETY CLASS 1. THIS UNIT MUST
BE EARTHED.