Page 17
Setting Precision SL100 model
To set generator to SL100 model, generator must be programmed fi rst! Once programmed
follow steps below:
1. Power up the generator.
2. Simultaneously press and hold three pushbuttons: DB USER PB, MB USER PB and MB
RESET PB.
3. While observing PSU On LED, release MB RESET PB whilst still holding DB USER PB and
MB USER PB. Hold both pushbuttons for about 5 seconds until PSU On LED blink once
indicating SL100 model has been set successfully.
To confi rm, power cycle generator while observing PSU On LED. One blink indicates
Precision SL100 and two blinks SL200 model.
Note: Alarm LED may blink/illuminate during above process, this is a normal operation.
For pushbutton position see fi gure 1 below.
Figure 1 – Precision SL PCB assembly Figure 1 – Precision SL PCB assembly