H² Detector
User Manual
Page 22
2013 © Peak Scientific – Rev 4 – 03/09/1
•
Contact Peak Service Provider.
F46 Cartridge Analog Failure –
Electronic Failure or gas
concentration greater than full
scale.
•
Replace Cartridge.
•
Contact Peak Service Provider.
F47 Cartridge Memory Invalid –
Checksum Error.
•
Replace Cartridge
•
Contact Peak Service Provider.
F48 Cartridge Absent – No
communications.
•
Reseat Cartridge.
•
Replace Cartridge.
•
Contact Peak Service Provider.
F49 Cartridge Wrong Type –
Cartridge type found to be
incorrect after boot-up.
•
Replace Cartridge or press
to accept if correct.
•
Contact Peak Service Provider.
F80 Temperature Limits Exceeded
– Temperature is outside limits.
•
Check installation environment.
•
Contact Peak Service Provider.
F81 Flow Fail – Flow <70% of
nominal for 15 seconds.
•
Check Filters.
•
Check for kinked tubing.
•
Replace Pump.
•
Contact Peak Service Provider.
F84 Misc. Transmitter Fault –
Transmitter is defective.
•
Service or replace Detector.
•
Contact Peak Service Provider.