4.9
View
When entering "VIEW" menu, an operator is offered to view and analyze current testing results
as well as the saved results. In order to view the current testing results, it is necessary to press
key.
Fig. 24
– "VIEW" menu
Note. To view previously saved testing results, it is necessary to press
key, choose
the "RESULTS" submenu and press
key. Then, move the cursor to the certain result and
press
key
– "F1 Load". Then, an automatic enter in "VIEW" is carried out.
Position “Position” – moves the measurement cursor in the time chart.
Position “Width” – sets the width of the measuring cursor.
Position “Type” – the type of cursor measurement. Four types of measurement are available.
° Pk-Pk
– the measurements are made between two points situated at the max. distance from
each other in cursor width.
° Cnt-Pk - the measurements are made between the cursor centre and the point situated at the
max. distance from the centre in cursor width.
° Cnt-K - the measurements are made between the centre and the cross point of cursor centre
signal.
° Vrt-M
– vertical maximum - the measurements are made between two points situated at the
max. distance from each other in the cursor width in vertical projection.
Position “Measurer” – is intended for meter signal parameters displaying in complex plane.
Position “Scan” – is for changing the number of points in a strip chart. When moving the cursor
in the strip chart, the complex plane display is updated with the signal, which is located within
the cursor width. The values of amplitude in volts, phase in degrees and defect depth in
millimetres or percent are displayed in the information line.