background image

Wireless Optical Dissolved Oxygen Sensor

Theory of Sensor Operation

8

013-16057A

The reference is a beam of red light emitted by a LED in the probe that reflects from the dye layer back to the 
photodiode. The lifetime of the luminescence from excitation by the blue light is compared to that of the reference 
value (red light) and a dissolved oxygen concentration is calculated by the sensor.

Optical and electrochemical dissolved oxygen probes determine the partial pressure of oxygen dissolved in the 
sample being tested. The dissolved oxygen partial pressure is converted to concentration (mg/l) or saturation 
percentage. The conversion from dissolved oxygen pressure to saturation percentage is done by dividing the 
reading by 160, the pressure of oxygen in air at 760 millimeters (mm) of mercury (Hg) and then multiplying by 
100%. For example, a measured dissolved oxygen pressure reading of 150 mm Hg would be converted to 93.8% 
saturation (150/160 x 100%).

See the PASCO web site at www.pasco.com for more information.

Factors that Affect Dissolved Oxygen Measurement

Factors that affect dissolved oxygen measurement include temperature, barometric pressure, and salinity. 

Temperature: The rate of diffusion of oxygen through the Sensor Cap changes with temperature (approximately 
1.5% per degree celsius). The Optical Dissolved Oxygen Probe has a built-in thermistor, and the temperature 
changes are compensated for by proprietary algorithms that use readings from the thermistor.

Barometric Pressure: The partial pressure of oxygen in a sample of air or water is not affected by barometric 
pressure, but dissolved oxygen concentration in milligrams per liter is affected. The Optical Dissolved Oxygen 
Probe has a built-in barometer sensor, and the reading from that device is used to determine the oxygen pressure 
during calibration. 

Salinity: As the salinity of water increases, its ability to dissolve oxygen decreases. (Note: The user will be able to 
enter a value for the “salinity constant” in the data acquisition software.) For precise measurement of salinity, the 
PS-2195 Salinity Sensor is recommended (see the web site at www.pasco.com).

Water Type

Average Salinity

Fresh water

<0.5 ppt*

Brackish water

0.5 to 30 ppt

Sea water

33 to 47 ppt

Saline water

30 to 50 ppt

Brine

>50 ppt

Optical Dissolved Oxygen 

Probe

Sensing Layer (dye layer)

Sensor Cap with 

Diffusion Layer

Photodiode and Light 

Emitting Diodes (LEDs)

DO NOT TOUVH

Rubber O-ring

NOTE: Avoid touching 

the Diffusion Layer or 

the Sensing Layer

Summary of Contents for PS-3224

Page 1: ...ommended Equipment Optical Dissolved Oxygen Metal Guard PS 3605 Required Item Part Number PASCO Data Collection Software see www pasco com Battery LED Bluetooth LED Micro USB Cable Transparent Cover E...

Page 2: ...ensor including the following Barometric pressure is recorded and stored only when the sensor is initially connected via Bluetooth or USB Initial Step Charge the Battery Connect the Cable Remove the t...

Page 3: ...In Sparkvue open the Main Menu and select Help for information about collecting displaying and analyzing data In the window that opens is the address of the web page for Help Copy and paste the addre...

Page 4: ...vue and PASCO Capstone support logging Check the PASCO Web page at www pasco com software for the latest software version Set Up the Software SPARKvue Set Up Connecting the Sensor to a Tablet or a Com...

Page 5: ...sor to a Computer with the Micro USB Cable Connect the micro end of the included Micro USB Cable into the micro USB port on the top of the sensor Connect the other end of the Micro USB Cable to a USB...

Page 6: ...issolved Oxygen Sensor Set Up the Software 6 013 16057A Select a Layout and then select a data display In the Data Display Click Select Measurement Select a measurement from the list under Sensors Sel...

Page 7: ...communicating with the computer software or tablet application try restarting the software or application If the problem remains press and hold the ON button on the sensor for 10 seconds and then rel...

Page 8: ...c pressure and salinity Temperature The rate of diffusion of oxygen through the Sensor Cap changes with temperature approximately 1 5 per degree celsius The Optical Dissolved Oxygen Probe has a built...

Page 9: ...int free cloth If necessary use a mild detergent Do not use alcohols or other organic solvents that may deteriorate the oxygen permeable diffusion layer and the sensing layer dye layer When changing t...

Page 10: ...ure of the gas in wet air or in solution is taken to be 20 9 Given this it is reasonable to use the probe in a gaseous environment to measure the oxygen present The sensor outputs a value labeled O2 G...

Page 11: ...ssolve enough sodium sulfite sodium bisulfite or potassium sulfite in 150 milliliters of distilled water such that a small amount of sodium sulfite remains undissolved in the bottom of the beaker WARN...

Page 12: ...connect pair and calibrate the other sensor s if necessary During the calibration process if you click Calibrate first point or Calibrate second point and then click Cancel the sensor will be restored...

Page 13: ...RKvue Calibration Steps Place the Wireless Optical Dissolved Oxygen Sensor into your known standard for 100 dissolved oxygen saturation 1 In SPARKvue touch or click Sensor Data in the Welcome Screen a...

Page 14: ...rdware Setup icon to open the Configure Sensors and Measurements window 5 In the Configure sensors and measurements window D02 Saturation is the default choice for calibration Touch or Click the Calib...

Page 15: ...the Calibrate Sensor Enter Values window Calibration Point 2 is the default selection with the Standard Value 100 0000 and the Sensor Value panel NOTE that the Sensor Value panel will show the number...

Page 16: ...ight arrow in the Configure sensors and measurements panel to close the panel and return to the data dis play 2 In the data display click the up arrow menu icon in the Live Data Bar to open the menu t...

Page 17: ...ing caused by exposure to light so the Sensor Cap must be replaced periodically It should also be replaced if it is cracked scratched or damaged The Sensor Cap is warranted for one year and may last m...

Page 18: ...Cap Calibration After a new Replacement Sensor Cap PS 3604 is placed on the Optical Dissolved Oxygen probe there are two parts to the Sensor Cap Calibration process First enter the Sensor Calibration...

Page 19: ...ace between each number NOTE Enter the numbers exactly as they appear on the sheet with the Calibration Code Coefficients for Cap Replacement including negative signs decimal points and other characte...

Page 20: ...n numbers from the sheet with the Calibration Code Coefficients for Cap Replacement as a list of 10 numbers separated by spaces NOTE Enter the numbers with a space between each number NOTE Enter the n...

Page 21: ...r calibration above 200 10 Resolution 0 01 mg l or 0 1 saturation Temperature C Range 0 50 Accuracy 0 1 C Resolution 0 1 C Barometric Pressure inHg Optional Units mmHg Range 375 836 mmHg Accuracy TBD...

Page 22: ...not be disposed of in a standard waste container Battery Disposal Instructions Batteries contain chemicals that if released may affect the environment and human health Batteries should be collected se...

Page 23: ...ng lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The Ma...

Page 24: ...Wireless Optical Dissolved Oxygen Sensor Appendix A 24 013 16057A...

Reviews: