![PASCO PS-3224 Reference Manual Download Page 18](http://html2.mh-extra.com/html/pasco/ps-3224/ps-3224_reference-manual_3876816018.webp)
Wireless Optical Dissolved Oxygen Sensor Replacement Sensor Cap Calibra-
18
013-16057A
Replace the Sensor Cap
To remove the Sensor Cap, first remove the Rubber Boot from the bottom of the probe. Grasp the plastic section of
the Probe body in one hand and turn the metal Sensor Cap counterclockwise to loosen it.
DO NOT USE A TOOL
to unscrew the Sensor Cap. If the Rubber O-ring is damaged, replace it with the O-ring included with the
Replacement Sensor Cap. Put the Replacement Sensor Cap onto the end of the probe and turn it clockwise until it
meets the metal part of the probe body.
DO NOT USE A TOOL
to install the new Sensor Cap.
NOTE
: The Sensor
Cap may be quite difficult to remove and replace. Make sure that your hands, the probe and the sensor cap are dry
and clean.
Replacement Sensor Cap Calibration
After a new Replacement Sensor Cap (PS-3604) is placed on the Optical Dissolved Oxygen probe, there are two
parts to the Sensor Cap Calibration process. First, enter the Sensor Calibration Coefficients into the software
(PASCO Capstone or SPARKvue). Second, do a one-point or two-point calibration as described previously.
•
The Replacement Sensor Cap includes a page with the Sensor Calibration Coefficients for Cap Replacement
(please see the example page).
Optical Dissolved Oxygen
Probe
Sensor Cap
Rubber O-ring
NOTE: DO NOT
TOUCH
Metal Probe
Body
Sensor Calibration Coefficients for Cap Replacement
SN?YL0118072007
Sensor Calibration Coefficients
K0?-0.0002
K1?0.2814
K2: 0.0423
K3?-0.0015
K4?0
K5?0.0133997
K6?0.0000903
K7?30.1
K8?1.881343
K9?BA8D
Please keep it for future reference
Example Numbers:
Do not use these
numbers for
calibration.
SERIAL NUMBER for the
Coefficients.
Use it as a reference if you
contact Technical Support
Example page
of Sensor
Calibration
Coefficients