background image

Wireless Optical Dissolved Oxygen Sensor

Calibration: See www.pas-

10

013-16057A

Using the Wireless Optical Dissolved Oxygen Sensor Submerged

Typically, the probe of the Wireless Optical Dissolved Oxygen Sensor would sit 
in a beaker of the liquid sample that is being measured. Alternately, the sensor 
could be held by a clamp so that the end of the probe is immersed in the 
container with the liquid sample. In addition, the sensor is designed to be 
submerged for such uses as monitoring water quality in a pond or stream. Turn 
the sensor on and pair it with the computing device. Set the sensor so that it will 
record in logging mode. (The sensor will not connect wirelessly when 
submerged.) To protect the sensor from the liquid sample, screw the transparent 
cover firmly onto the top of the sensor. Make sure that the bottom edge of the 
transparent cover touches the narrow lip. Remove but do not discard the Rubber 
Boot before making a measurement. The sensor is not buoyant so it will sink if 
dropped into the liquid sample. Therefore, attach a cable or cord to the eyebolt 
at the top of the transparent cover before submerging the sensor. Be careful to 
hold the cable or cord securely so that the sensor can be retrieved.

Using the Wireless Optical Dissolved Oxygen Sensor in Air 

The Wireless Optical Dissolved Oxygen Sensor probe was specifically designed to be used in water. However, 
reasonably good qualitative results may be obtained in air. 

The optical dissolved oxygen probes are generally calibrated in 100% air-saturated water or in 'wet' air that is in a 
100% relative humidity environment. In either case, the oxygen contribution to the partial pressure of the gas in 
'wet' air or in solution is taken to be 20.9%. Given this, it is reasonable to use the probe in a gaseous environment 
to measure the oxygen present. The sensor outputs a value labeled “O2 Gas Concentration”. 

This value is calculated by taking the DO2 Saturation value from the sensor (100% in air-saturated water or 'wet' 
air) and multiplying it by 0.209(20.9%), the assumed partial pressure contribution of O2 in air. For example, if the 
DO2 Saturation value from the sensor is 80%, the calculated partial pressure contribution of O2 in air is 80% X 
0.209 = 16.7% O2.

Best accuracy will be achieved in high humidity environments. Long term measurements in dry air are not 
recommended as this will dry out the probe Sensor Cap and may affect future measurements with the probe. 

NOTE: The probe should not be used in an environment with flammable, caustic or corrosive gases. 

Accessory: Optical Dissolved Oxygen 
Probe Guard

 (PS-3605)

The Wireless Optical Dissolved Oxygen Probe Guard 
(PS-3605) is designed to thread onto the probe to 
protect the Sensor Cap. It also enables the probe to 
sink more rapidly when the probe is immersed in 
water. 

NOTE: Do not use any tool to thread the Probe Guard 
onto the probe. Avoid touching the end of the Sensor 
Cap at all times.

Calibration: See 
www.pasco.com/calibration

The Wireless Optical Dissolved Oxygen Sensor is calibrated at the factory and does not need to be re-calibrated 
under most circumstances. However, if the Sensor Cap is replaced with a new Sensor Cap (PS-3604), then the 
Sensor Cap Calibration Code Coefficients (included with the new Sensor Cap) will need to be entered into the 
sensor using the PASCO data collection software (PASCO Capstone or SPARKvue). The sensor will then need to 
be re-calibrated. (See Replacement Sensor Cap Calibration.)

Narrow 

Lip

Transparent 

Cover

Wireless Optical Dissolved 

Oxygen Probe Guard

Summary of Contents for PS-3224

Page 1: ...ommended Equipment Optical Dissolved Oxygen Metal Guard PS 3605 Required Item Part Number PASCO Data Collection Software see www pasco com Battery LED Bluetooth LED Micro USB Cable Transparent Cover E...

Page 2: ...ensor including the following Barometric pressure is recorded and stored only when the sensor is initially connected via Bluetooth or USB Initial Step Charge the Battery Connect the Cable Remove the t...

Page 3: ...In Sparkvue open the Main Menu and select Help for information about collecting displaying and analyzing data In the window that opens is the address of the web page for Help Copy and paste the addre...

Page 4: ...vue and PASCO Capstone support logging Check the PASCO Web page at www pasco com software for the latest software version Set Up the Software SPARKvue Set Up Connecting the Sensor to a Tablet or a Com...

Page 5: ...sor to a Computer with the Micro USB Cable Connect the micro end of the included Micro USB Cable into the micro USB port on the top of the sensor Connect the other end of the Micro USB Cable to a USB...

Page 6: ...issolved Oxygen Sensor Set Up the Software 6 013 16057A Select a Layout and then select a data display In the Data Display Click Select Measurement Select a measurement from the list under Sensors Sel...

Page 7: ...communicating with the computer software or tablet application try restarting the software or application If the problem remains press and hold the ON button on the sensor for 10 seconds and then rel...

Page 8: ...c pressure and salinity Temperature The rate of diffusion of oxygen through the Sensor Cap changes with temperature approximately 1 5 per degree celsius The Optical Dissolved Oxygen Probe has a built...

Page 9: ...int free cloth If necessary use a mild detergent Do not use alcohols or other organic solvents that may deteriorate the oxygen permeable diffusion layer and the sensing layer dye layer When changing t...

Page 10: ...ure of the gas in wet air or in solution is taken to be 20 9 Given this it is reasonable to use the probe in a gaseous environment to measure the oxygen present The sensor outputs a value labeled O2 G...

Page 11: ...ssolve enough sodium sulfite sodium bisulfite or potassium sulfite in 150 milliliters of distilled water such that a small amount of sodium sulfite remains undissolved in the bottom of the beaker WARN...

Page 12: ...connect pair and calibrate the other sensor s if necessary During the calibration process if you click Calibrate first point or Calibrate second point and then click Cancel the sensor will be restored...

Page 13: ...RKvue Calibration Steps Place the Wireless Optical Dissolved Oxygen Sensor into your known standard for 100 dissolved oxygen saturation 1 In SPARKvue touch or click Sensor Data in the Welcome Screen a...

Page 14: ...rdware Setup icon to open the Configure Sensors and Measurements window 5 In the Configure sensors and measurements window D02 Saturation is the default choice for calibration Touch or Click the Calib...

Page 15: ...the Calibrate Sensor Enter Values window Calibration Point 2 is the default selection with the Standard Value 100 0000 and the Sensor Value panel NOTE that the Sensor Value panel will show the number...

Page 16: ...ight arrow in the Configure sensors and measurements panel to close the panel and return to the data dis play 2 In the data display click the up arrow menu icon in the Live Data Bar to open the menu t...

Page 17: ...ing caused by exposure to light so the Sensor Cap must be replaced periodically It should also be replaced if it is cracked scratched or damaged The Sensor Cap is warranted for one year and may last m...

Page 18: ...Cap Calibration After a new Replacement Sensor Cap PS 3604 is placed on the Optical Dissolved Oxygen probe there are two parts to the Sensor Cap Calibration process First enter the Sensor Calibration...

Page 19: ...ace between each number NOTE Enter the numbers exactly as they appear on the sheet with the Calibration Code Coefficients for Cap Replacement including negative signs decimal points and other characte...

Page 20: ...n numbers from the sheet with the Calibration Code Coefficients for Cap Replacement as a list of 10 numbers separated by spaces NOTE Enter the numbers with a space between each number NOTE Enter the n...

Page 21: ...r calibration above 200 10 Resolution 0 01 mg l or 0 1 saturation Temperature C Range 0 50 Accuracy 0 1 C Resolution 0 1 C Barometric Pressure inHg Optional Units mmHg Range 375 836 mmHg Accuracy TBD...

Page 22: ...not be disposed of in a standard waste container Battery Disposal Instructions Batteries contain chemicals that if released may affect the environment and human health Batteries should be collected se...

Page 23: ...ng lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The Ma...

Page 24: ...Wireless Optical Dissolved Oxygen Sensor Appendix A 24 013 16057A...

Reviews: