background image

Wireless Temperature Link

Technical Support

8

013-16067A

In the software, turn Bluetooth off and then turn it back on. Retry.

Temperature Probe Maintenance

Rinse and dry the temperature probe before putting the sensor away for storage. 

Suggested Experiments

Practically any experiment that uses a temperature measurement can be done with the PS-3222 Wireless 
Temperature Link and the Fast Response Temperature Probe. In addition, the Wireless Temperature Link is 
designed to work with other PASCO temperature probes and with the TD-8595 Absolute Zero Sphere and the 
TD-8596A Ideal Gas Law Apparatus.

See the PASCO Web site at www.pasco.com/products/lab-manuals for more information about experiments.

Sensor Specifications

Technical Support

For assistance with any PASCO product, contact PASCO at:

Check the PASCO website for the latest version of the instruction manual.

www.pasco.com/manuals

Limited Warranty 

For a description of the product warranty, see the PASCO catalog. 

Copyright 

The PASCO scientific 

Instruction Manual

 is 

copyrighted with all rights reserved. Permission is granted to non-profit educational institutions for reproduction of any part of this manual, 
providing the reproductions are used only in their laboratories and classrooms, and are not sold for profit. Reproduction under any other 
circumstances, without the written consent of PASCO scientific, is prohibited. Rev: 02/19. 

Trademarks 

PASCO, PASCO Capstone, and 

SPARKvue are trademarks or registered trademarks of PASCO scientific, in the United States and/or in other countries. For more information 
visit 

www.pasco.com/legal. 

Product End of Life Disposal Instructions:

This electronic product is subject to disposal and recycling regulations that vary by country and region. It is your responsibility to recycle your 
electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects human health 
and the environment. To find out where you can drop off your waste equipment for recycling, please contact your local waste recycle/disposal 
service, or the place where you purchased the product.

Wireless Temperature 

Link

Range

-

35

°C to +1

3

5°C 

Accuracy

±0.5° C

Resolution

0.0

5

°C

Sample Rate

Up to 

2

0 Hz

Address: PASCO scientific

10101 Foothills Blvd.
Roseville, CA 95747-7100

Web:

www.pasco.com

Phone: +1 916-462-8384 (worldwide)

877-373-0300 (U.S.)

Email

[email protected]

Summary of Contents for PS-3222

Page 1: ...o com Compatible Equipment Temperature Range PS 2131 PASPORT Skin Surface Temperature Probe 35 to 135 C PS 2133 Stainless Steel Temperature Probe 35 to 135 C TD 8595 Absolute Zero Sphere 35 to 135 C T...

Page 2: ...ded and is well suited for continuous recording and discrete measurements The sensor is designed to optimize the battery usage time Since each sensor has a unique Device ID number more than one sensor...

Page 3: ...nnected and after several minutes if not connected Set up Software Help In Sparkvue open the Main Menu and select Help for information about collecting displaying and analyzing data In the window that...

Page 4: ...Configuration screen look in the Connected Devices list Touch or click to select the correct address that matches the Device ID XXX XXX number found on the Wireless Temperature Link A message confirms...

Page 5: ...nect the Sensor in PASCO Capstone 1 Turn on the sensor and start the software In PASCO Capstone click or touch the Hardware Setup icon in the Tools palette to open the Hard ware Setup panel The Hardwa...

Page 6: ...ing the Wireless Temperature Link Calibration is not always necessary especially if you are measuring a change in temperature rather than absolute temperature values However it is possible to calibrat...

Page 7: ...abs on the cover The side of the battery with the on it should be against the cover Replace the Battery Compartment Cover Put the Battery Compartment Door with the new battery back onto the sensor Ali...

Page 8: ...rofit educational institutions for reproduction of any part of this manual providing the reproductions are used only in their laboratories and classrooms and are not sold for profit Reproduction under...

Page 9: ...to a USB port allows up to three Bluetooth SMART devices such as this PASCO wireless device to connect to some Windows computers some Chromebooks and older Macintosh computers Note The PS 3500 USB Blu...

Page 10: ...mputer as explained earlier 2 In Capstone click the Calibration icon in the Tools palette to open the Calibration window In Step 1 select the Temperature measurement and click Next In Step 2 the Wirel...

Page 11: ...te Sensor Select Measurement window In this window the sensor measurement and calibration type are already selected 5 Click or touch Continue to open the Calibrate Sensor Enter Values window 6 Put the...

Page 12: ...Capstone is to take the stream of raw data from a sensor and transform it into the calibrated data that you see in the Graph Table and other displays If you do not calibrate a sensor yourself the sof...

Page 13: ...l automatically select the most typical calibration type as the default setting Two Point In a two point calibration you reset two points to define a new line This type of calibration affects both the...

Page 14: ...Wireless Temperature Link Appendix B Calibration 14 013 16067A...

Reviews: