PASCO PS-3222 Reference Manual Download Page 10

Wireless Temperature Link

Appendix B: Calibration

10

013-16067A

1

The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011. The MacBook Pro was 

updated in 2012. The Mac Pro that debuted in December 2013 has Bluetooth SMART support.

Exception

: Before you upgrade to El Capitan (Mac OS X 10.11.x), if you have a Macintosh with LMP version “0x4” 

that requires the PS-3500 USB Bluetooth 4.0 Adapter, please contact PASCO Technical Support for further 
instructions.

Appendix B: Calibration

Calibration is not always necessary, especially if you are measuring a change in temperature rather than absolute 
temperature values. However, it is possible to calibrate the sensor. 

Prepare for Calibration

Calibration will need an ice-water bath, a container of hot water, and a thermometer. The sensor will need to be 
“paired” with a tablet or computer, and the data collection software (for example, SPARKvue or Capstone) should 
be running.

Capstone Calibration 

1.

Turn on the sensor. Start Capstone. “Pair” the Wireless 
Temperature Link wirelessly to the computing device (e.g., 
computer) as explained earlier.

2.

In Capstone, click the “Calibration” icon in the 
Tools palette to open the “Calibration” window. 
In Step 1, select the Temperature measurement 
and click “Next”. In Step 2 the “Wireless 
Temperature Link” is picked by default. In Step 
3, Two Standards (2 point) is the default choice. 
Click or touch “Next”. 

3.

In Step 4, put the thermometer into the ice-water 
bath. Also put the end of the Fast Response 
Temperature Probe in the ice-water bath. When 
the “Current Value” stabilizes, enter the 
temperature reading in the “Standard Value” text 
box. Click “Set Current Value to Standard 
Value”. Click “Next” to open Step 5.

4.

In Step 5, move the thermometer and the Fast 
Response Temperature Probe to the container 
of hot water. When the “Current Value” 
stabilizes, enter the temperature reading in the 
“Standard Value” text box. Click “Set Current 
Value to Standard Value”. Click “Next”. Review 
the calibration and then click “Finish”. Close the 
Calibration window.

SPARKvue Calibration

1.

Turn on the sensor and start SPARKvue.

2.

In the “Welcome” screen, click or touch “Sensor Data” to open the “Sensor Data Configuration” screen. 
Connect (pair) the Wireless Temperature Link under “Connected Devices” as described earlier. 

Highlight the 

number and 

then enter 

the exact 

volume here.

Next, click 

here.

Summary of Contents for PS-3222

Page 1: ...o com Compatible Equipment Temperature Range PS 2131 PASPORT Skin Surface Temperature Probe 35 to 135 C PS 2133 Stainless Steel Temperature Probe 35 to 135 C TD 8595 Absolute Zero Sphere 35 to 135 C T...

Page 2: ...ded and is well suited for continuous recording and discrete measurements The sensor is designed to optimize the battery usage time Since each sensor has a unique Device ID number more than one sensor...

Page 3: ...nnected and after several minutes if not connected Set up Software Help In Sparkvue open the Main Menu and select Help for information about collecting displaying and analyzing data In the window that...

Page 4: ...Configuration screen look in the Connected Devices list Touch or click to select the correct address that matches the Device ID XXX XXX number found on the Wireless Temperature Link A message confirms...

Page 5: ...nect the Sensor in PASCO Capstone 1 Turn on the sensor and start the software In PASCO Capstone click or touch the Hardware Setup icon in the Tools palette to open the Hard ware Setup panel The Hardwa...

Page 6: ...ing the Wireless Temperature Link Calibration is not always necessary especially if you are measuring a change in temperature rather than absolute temperature values However it is possible to calibrat...

Page 7: ...abs on the cover The side of the battery with the on it should be against the cover Replace the Battery Compartment Cover Put the Battery Compartment Door with the new battery back onto the sensor Ali...

Page 8: ...rofit educational institutions for reproduction of any part of this manual providing the reproductions are used only in their laboratories and classrooms and are not sold for profit Reproduction under...

Page 9: ...to a USB port allows up to three Bluetooth SMART devices such as this PASCO wireless device to connect to some Windows computers some Chromebooks and older Macintosh computers Note The PS 3500 USB Blu...

Page 10: ...mputer as explained earlier 2 In Capstone click the Calibration icon in the Tools palette to open the Calibration window In Step 1 select the Temperature measurement and click Next In Step 2 the Wirel...

Page 11: ...te Sensor Select Measurement window In this window the sensor measurement and calibration type are already selected 5 Click or touch Continue to open the Calibrate Sensor Enter Values window 6 Put the...

Page 12: ...Capstone is to take the stream of raw data from a sensor and transform it into the calibrated data that you see in the Graph Table and other displays If you do not calibrate a sensor yourself the sof...

Page 13: ...l automatically select the most typical calibration type as the default setting Two Point In a two point calibration you reset two points to define a new line This type of calibration affects both the...

Page 14: ...Wireless Temperature Link Appendix B Calibration 14 013 16067A...

Reviews: