PS-3220

 Set-up

7

013-16067A

the old battery, installing a new battery of the same type, and replacing the Battery Compartment Cover. The 
procedure needs a coin and a CR2032 three volt battery. 

(NOTE: This battery is available from PASCO and in electronic and commercial stores.)

Remove the Battery Compartment Cover

Hold the sensor bottom-side up. Use a coin in the slot to turn the Battery 
Compartment Cover counterclockwise (left-to-right) until the indicator on the 
door is aligned with the second mark on the sensor. 

Turn the sensor bottom-side down so that the Battery Compartment Cover can 
drop into the palm of your hand. Make sure that the battery cover gasket stays 
on the cover. The gasket is an “O-ring” that is held in place by the tabs on the 
cover. Please do not touch the inside of the battery compartment

Remove the used battery from the Battery Compartment Cover and replace it with a new identical type battery. 
Note that the battery is held in place by the small tabs on the cover. The side of the battery with the “+” on it 
should be against the cover.

Replace the Battery Compartment Cover

Put the Battery Compartment Door with the new battery back onto the sensor. Align the 
indicator on the cover with the second mark, and use the coin in the slot to turn the 
cover clockwise (right-to-left) until the indicator is aligned with the first mark on the 
sensor.

(See “Battery Disposal Instructions” under Technical Support.)

Related Item

Coin-cell Battery Replacement Pack (10 pack) - PS-3504

See the PASCO web site at www.pasco.com for more information.

Troubleshooting the Wireless Temperature Link

If the sensor loses Bluetooth connection and will not reconnect, try cycling the ON button. Press and briefly 
hold the button until the status LEDs blink, and then release the button 

If the sensor stops communicating with the computer software or tablet application, try restarting the software 
or application. If the problem remains, press and hold the ON button for ten seconds and then release the 
button. Turn on the sensor in the usual way.

Second mark

Coin 

slot

Battery 

Compartment 

Cove

Gasket (“O”-ring)

Battery (CR2032)

Tab

PLEASE Do Not Touch

First mark

Summary of Contents for PS-3222

Page 1: ...o com Compatible Equipment Temperature Range PS 2131 PASPORT Skin Surface Temperature Probe 35 to 135 C PS 2133 Stainless Steel Temperature Probe 35 to 135 C TD 8595 Absolute Zero Sphere 35 to 135 C T...

Page 2: ...ded and is well suited for continuous recording and discrete measurements The sensor is designed to optimize the battery usage time Since each sensor has a unique Device ID number more than one sensor...

Page 3: ...nnected and after several minutes if not connected Set up Software Help In Sparkvue open the Main Menu and select Help for information about collecting displaying and analyzing data In the window that...

Page 4: ...Configuration screen look in the Connected Devices list Touch or click to select the correct address that matches the Device ID XXX XXX number found on the Wireless Temperature Link A message confirms...

Page 5: ...nect the Sensor in PASCO Capstone 1 Turn on the sensor and start the software In PASCO Capstone click or touch the Hardware Setup icon in the Tools palette to open the Hard ware Setup panel The Hardwa...

Page 6: ...ing the Wireless Temperature Link Calibration is not always necessary especially if you are measuring a change in temperature rather than absolute temperature values However it is possible to calibrat...

Page 7: ...abs on the cover The side of the battery with the on it should be against the cover Replace the Battery Compartment Cover Put the Battery Compartment Door with the new battery back onto the sensor Ali...

Page 8: ...rofit educational institutions for reproduction of any part of this manual providing the reproductions are used only in their laboratories and classrooms and are not sold for profit Reproduction under...

Page 9: ...to a USB port allows up to three Bluetooth SMART devices such as this PASCO wireless device to connect to some Windows computers some Chromebooks and older Macintosh computers Note The PS 3500 USB Blu...

Page 10: ...mputer as explained earlier 2 In Capstone click the Calibration icon in the Tools palette to open the Calibration window In Step 1 select the Temperature measurement and click Next In Step 2 the Wirel...

Page 11: ...te Sensor Select Measurement window In this window the sensor measurement and calibration type are already selected 5 Click or touch Continue to open the Calibrate Sensor Enter Values window 6 Put the...

Page 12: ...Capstone is to take the stream of raw data from a sensor and transform it into the calibrated data that you see in the Graph Table and other displays If you do not calibrate a sensor yourself the sof...

Page 13: ...l automatically select the most typical calibration type as the default setting Two Point In a two point calibration you reset two points to define a new line This type of calibration affects both the...

Page 14: ...Wireless Temperature Link Appendix B Calibration 14 013 16067A...

Reviews: