HPVA Series Operator’s Manual
Connecting Gases
Mar 2011
6-7
Required Items.
Particulate Systems Supplied
User Supplied
Stainless steel gas tubing
Two 7/16-in. wrenches
Analysis gas (recommended purity 99.999%)
Helium (recommended purity 99.999%)
Clean compressed air and air line
Regulators for gas bottles
Disconnecting the Depleted Bottle
1. Close the gas bottle shut-off valve, then open the regulator shut-off valve.
Gas Bottle
shut-off valve
Regulator
shut-off
valve
2. Both gauges should read at or near zero; if not, open the gas inlet valve (3) and the vent valve (2)
and allow the pressure to purge from the lines.
3. Use an appropriate wrench to loosen the nut at the regulator/gas bottle connection, then remove the
regulator from the bottle.
You do not have to disconnect the gas line from the regulator or the instrument.
4. Replace the protective cap on the depleted bottle, disconnect the retaining strap, and remove the
bottle from its current location.
Summary of Contents for HPVA Series
Page 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Page 4: ......
Page 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Page 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Page 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Page 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Page 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Page 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Page 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Page 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...