HPVA Series Operator’s Manual
Purging the System
Mar 2011
2-15
Purging the System
When the HPVA is performing an experiment or being operated in
manual control mode, the valve switches on the front panel should be
placed and remain in the closed (down) position.
If a switch is placed in the opened position during a software
operation, it will override the software and may damage the
instrument.
1. Click
Manual Control
from the Main Menu to display the Manual Control screen.
24.00
1.000
1.000
31.00
Manifold temperature
Manifold pressure
Sample pressure
Sample temperature
You can open and close a valve by double-clicking the left mouse button on the valve icon or
clicking the valve icon then pressing the space bar. Yellow indicates a closed valve and green indi-
cates an open valve.