Definitions of Fields in the System Windows
HPVA Series Operator’s Manual
2-30
Mar 2011
Ambient
Temp:
Enter the temperature of the air in the laboratory in °C.
Hold:
Enter the number of minutes the system will wait once the
sample RTD reaches the ambient temperature.
Evacuation Time
Enter the number of minutes you wish to evacuate the sample
prior to running the free space routine.
Adsorbate Name
Enter the name of the gas to be used for the experiment. It should
match the name under the
Adsorbate
column on the
Gases
tab of
the Excel macro spreadsheet.
Gas Port
Select the port (either 6 or 7) to which the adsorbate gas tank is
attached to the instrument.
If the adsorbate is a gas mixture, check
Mixture
.
Pressure Steps
This table lists the pressures the sample holder will reach during
the experiment when the standard dosing mode is used.
To enter a pressure step, enter the pressure in bars in the
Pressure
field, then click
Add
.
If you wish to delete a pressure step, click the step in the table,
then click
Delete
.
Data Logging Interval
This field enables you to specify how often data points are
collected either by time or by pressure.
Enter a number of minutes in the
min
field or a pressure in the
bars
field.
This is typically set to every 2 minutes or whenever the system
pressure changes by 0.005 bars.
Port 1 through Port 4
Click
Port 1
.
Measure FS
Check this box if you wish to have a free space analysis conducted
prior to the isotherm experiment.
Ambient FS
If the Measure FS box is not checked, enter the ambient free space
volume of the sample tube in this field.
Summary of Contents for HPVA Series
Page 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Page 4: ......
Page 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Page 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Page 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Page 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Page 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Page 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Page 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Page 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...