6. You can click the file name tab to display experiment data and the volume dosed and volume
adsorbed calculations.
The columns in the calculation tables are described below.
Column Heading
Description
ReqPm
Requested pressure by the user
Target Pm
Manifold target pressure for dosing
PmA
Pressure of manifold before dosing
TmA
Temperature of manifold before dosing
ZA
Compressibility of adsorbate at PmA and TmA
Vol A
Volume of adsorbate in manifold before dosing
PmB
Pressure of manifold after dosing
TmB
Temperature of manifold after dosing
ZB
Compressibility of adsorbate at PmB and TmB
Vol B
Volume of adsorbate in manifold after dosing