HPVA Series Operator’s Manual
Troubleshooting
Mar 2011
6-3
What Happened
Why
What to Do
Pressure cycling
(continued)
Gas is venting out of the
manifold too fast.
Adjust the
Vent
valve on the
instrument’s front panel.
Target pressure not
reached
Gas level in tank is too low.
Replace the gas bottle.
The pressure limit entered in
the Instrument Configuration
window does not match the
pressure set on the regulator.
For example, if you are using
CO
2
or methane, the pressure
limit must be set below the
maximum output of the
regulator.
Enter the correct pressure
limit.
Refer to
Ambient pressure not
reached
Slope set incorrectly in the
Hardware Configuration
window.
Enter the correct slope.
Degas or analysis not
progressing as it
should
O-rings in ports are missing,
worn, or damaged.
Replace the port O-ring.
Summary of Contents for HPVA Series
Page 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Page 4: ......
Page 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Page 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Page 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Page 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Page 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Page 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Page 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Page 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...