Lasair II Particle Counter Operations Manual
Page H-1
Appendix H
Pharmaceutical (Two-channel) Output
Mode
To support some pharmaceutical applications, the Lasair II particle counter can be
configured to display and print 0.5
μ
m and 5.0
μ
m
channels only.
LCD Screen Output
While the Pharmaceutical Two-channel mode is enabled, the
Data Table Screen
will display only 0.5
μ
m and 5.0
μ
m channel cumulative count data. Further, if “0.5
and 5.0 TCP/IP” is enabled, the data transmitted will only include these two
channels. Otherwise, all six channels of data will be transmitted.
Alarms
Alarms for all six channels are still available. It is advisable, however, to ensure that
only cumulative alarms are used, and that the alarms be only for the 0.5
μ
m and 5.0
μ
m channels. Differential alarms will not be valid for the derived two-channel data.
Enabling the Pharmaceutical Two-Channel Output Mode
This option is enabled from the
Comm Setup Screen
.
To enable the pharmaceutical two-channel output mode:
1
From the
Main Display Screen
, click the
Main Setup
icon softkey. The
Main Setup Screen will open.
2
Click the
Comm Setup
icon
softkey. The
Comm Setup Screen
will open.