TurbiTechw² LR Instruction Manual
9.5.2 Sample Result
Once the sample result has been established via the ‘Take Sample’ routine above, the result can be
entered into the sensor in the following way:
Press
to show the “MAIN MENU”.
Select “MEASUREMENT CONFIG” using
to highlight and press
.
Select measurement to be calibrated using
to highlight and press
.
Select “SAMPLE RESULT” using
to
highlight and press
.
Sensor details screen will be displayed,
press
to accept.
The time and date of the last Take
Sample will be displayed for the users
reference. Press
to accept.
The SAMPLE RESULT screen will be
displayed. This provides an opportunity
to enter the laboratory results from
previous take sample. Press
or
to
move the cursor below the digit to be
changed. Press
or
to increase or
decrease the digit. Enter the known value
of the sample.
Press
to accept the new calibration
value.
“Sample Result Successful” will now be
displayed: The sensor is now calibrated.
Press
several times to return back to the DISPLAY screen.
225769IM - Issue - 05 Issue Date 30/04/2019
Page 25 of 34
MEASUREMENT CONFIG
Measurement Status
Add Measurement
M:01 Turbidity (S:01)
M:01 CONFIG
M:01 Info
M:01 Set Zero
M:01 Averaging
M:01 Remove
M:01 Set Cal
M:01 Take Sample
M:01 Sample Result
M:01 SAMPLE RESULT
M:01 Turbidity
S:01 TurbiTechw
2
LR (0-30)
SN: 400001
Press OK to contiue or MENU to Exit
M:01 SAMPLE RESULT
Last Take Sample Details
Sample Date 25/03/2019 14:56:00
Sample User: System
Press OK to contiue or MENU to Exit
M:01 SAMPLE RESULT
3.40 NTU
Use to set value
Press OK to accept
Press MENU to cancel
M:01 SAMPLE RESULT
Applying Sample Result...
Please Wait
M:01 SAMPLE RESULT
Sample Result Successful
25/03/2019 15:00:08
Press OK or MENU to Exit
3.40
NTU