TurbiTechw² LR Instruction Manual
7 Sensor Configuration
Before attempting to configure the sensor, please read the user manual
that came with your monitor. The monitor manual will introduce you to
the basic set-up of the monitor, and will familiarise you with the monitor
menu structure and buttons.
The monitor leaves the factory with no sensors pre-installed.
Assuming the monitor has been physically connected to a sensor, the
next step is to register and configure the sensor before any measurements can be made. A single
sensor may provide one or more measurements. We advise only connecting one sensor at a time.
Once the first sensor has been registered, connect the second and register again. Repeat for any
additional sensors.
All sensors must be registered to the monitor in this way, even if they are different types.
Please note that live measurements are not available until the Sensor
Configuration stage has been completed.
7.1 Sensor Config
From the MAIN MENU screen, select SENSOR CONFIG by pressing
,
and press
to accept.
7.2 Sensor Status
This option allows the user to review the current status of the 8 sensor channels, these will all be set to
disabled until a sensor is added.
Once a sensor has been installed the display will be updated to indicate the sensor type installed and
it's status.
7.3 Add Sensor
1. From the MAIN MENU screen, select SENSOR CONFIG by
pressing
and press
.
2. The SENSOR MENU should be displayed. Press
to
highlight ADD SENSOR, and press
.
3. The Monitor will now search all possible addresses (0 to 240)
to find any attached sensors. During the search, any sensors found will be displayed momentary
before continuing with the search.
4. Once the search is complete, the Monitor will display a list of sensors found. Each sensor will be
automatically allocated a new address from S:01 to S:08.
5. Repeat the above process to install a second, third or more sensors. A total of 8 sensors are
possible (expansion box may be required to add additional sensors).
6. Sensor addition is now complete.
7. If a single or multiple sensors have been found the
can be pressed to escape from
continuing the search.
Page 16 of 34
225769IM - Issue - 05 Issue Date 30/04/2019
M:01 NTU
0.14
TURBIDITY
Monitor: OK 09:59:45 17/05/12
MAIN MENU
Monitor Config
Expansion Config
Sensor Config
Measurement Config
Alarm Config
Output Config
Information
SENSOR CONFIG
Sensor Status
Add Sensor
<No Sensors Installed>