SECTION 2 – PRODUCT SPECIFICATIONS 

 

Model Number 

MPC95 

MPC02 

MPC05 

MPC20 

Control Valve Type 

Normally closed proportional solenoid valve 

Maximum Flow Capacity (N

2

 unless otherwise 

noted) (Note 1) 

0.5 SLPM 

2.0 SLPM (N

2

 or Helium) 

5.0 SLPM 

20 SLPM 

Compatible Gases 

Nitrogen/air, oxygen, argon, carbon dioxide (all models except Model MPC02-BBNHP1); 

Helium (Model MPC02-BBNHP1 only) - Gas must be dry, clean and oil-free. 

Rangeability (Control Range) 
(Refer to Table 1) 

25:1 (4-100% full scale [FS]) 

50:1 (2-100% FS) 

Response Time 

1.0 second to within ±2% FS of setpoint (typical) 

Accuracy 

±2% FS (at 20°C and 30 PSIG) 

Repeatability 

±1% FS 

Temperature Coefficient 

±0.1% FS/°C (±0.056% FS/°F) 

Flow 

40% FS 

0.7% FS 

0.4% FS 

0.2% FS 

Flow 

10% FS 

Flow <40% FS 

1.2% FS 

0.7% FS 

0.3% FS 

Control 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Pressure 
Coefficient (Per 
14.5 PSI) 

Flow <10% FS 

2% FS 

1.2% FS 

0.5% FS 

0.2% FS 

Minimum Differential Pressure 
(Note 3) 

7 PSIG 

7 PSIG 

14.5 PSIG 

22 PSIG 

Maximum Differential Pressure 
(Note 4) 

40 PSIG 

Calibration Pressure (Note 2) 

30 PSIG (Inlet pressure: 30 PSIG and outlet pressure: 0 PSIG) 

Pressure 
 
 
 
 
 
 

Maximum Operating Pressure 

75 PSIG 

Calibration Temperature (Note 2) 

20°C 

Operating Temperature Range 

-10 to 50°C (14 to 122°F) 

Temperature 
 
 
 
 

Storage Temperature Range 

-10 to 60°C (14 to 140°F) 

Humidity  

Operating Humidity Range 

10 to 90% Relative Humidity (non-condensing) 

Setpoint Input 

Keypad Operation or External Setpoint Voltage Input 

Resolution 

Refer to Table 1 

Setpoint 
 
 
 
 

Setpoint Input Voltage 

0 to 5 Vdc or 1 to 5 Vdc (selectable) 

Display Type 

7-segment LED; 8 digits (Instantaneous flow rate display: 4 digits; Setpoint flow rate display: 4 digits) 

Display Resolution 

Refer to Table 1 

Flow Rate 
Indication 
 
 
 
 

Indication Accuracy 

±2% FS ±1 digit 

Display Range 

0.00 to 999,999.99L 

0.0 to 9,999,999.9L 

0.0 to 9,999,999.9L 

0 to 99,999,999L 

Display Resolution 

0.01L 

0.1L 

0.1L 

1L 

Every 5L count 

Every 20L count 

Every 50L count 

Every 200L count 

Totalizer 
Function 

Totalizer Backup Timing 

Every hour (time) from the previous backup 

Output Scale 

0 to full scale flow rate (scaling selectable) 

Output Signal Voltage 

0 to 5 Vdc or 1 to 5 Vdc (selectable) 

Maximum Output Signal Voltage 

7 Vdc maximum (maximum output signal when flow rate exceeds maximum flow capacity) 

Flow Rate 
Output 

Accuracy 

±0.5% FS (Input impedance of the connected device must be 100k ohms or greater). 

Overall output accuracy: Indication accuracy 

±

0.5% FS 

Number of Outputs 

Output Rating 

30 Vdc, 15 mAdc maximum (Open collector non-isolated output) 

Totalizer Pulse Output Width 

100 ms (

±

10%) (when totalizer pulse output is selected) 

Event Output 
 
 
 
 
 
 

Totalizer Pulse Output Rate 

0.01L/pulse 

0.1L/pulse 

0.1L/pulse 

1L/pulse 

Summary of Contents for Porter MPC Series

Page 1: ...MPC SERIES PANEL MOUNT MASS FLOW CONTROLLER INSTALLATION PROGRAMMING USER S MANUAL FM 1038 Rev 0 4 08...

Page 2: ...G 9 Location 9 Dimensional Data 9 Mounting Procedure 9 External Electrical Connections 11 Section 5 BASIC OPERATION 13 Switching Displays 13 Setting the Flow Rate 15 Selecting the Operation Mode 17 Se...

Page 3: ...positive shut off is a system requirement install a separate shut off valve When a separate shut off valve is closed the user must command the MFC control valve fully closed by one of the following me...

Page 4: ...User s Manual FM 1032 One 1 FEATURES The MPC Series Panel Mount MFC is a mass flow controller featuring high performance in a compact package design Integrating the advanced technologies of an ultra...

Page 5: ...rt stop reset can be remote controlled by external switching input After a reset by key operation the totalization calculation is started automatically However if a reset contact ON is carried out by...

Page 6: ...parate shut off valve AUTOMATIC RESET OF TOTALIZATION COUNT AT START OF CONTROL FUNCTION Start of control and totalization count reset can be carried out simultaneously by a single action key operatio...

Page 7: ...0 C 14 to 140 F Humidity Operating Humidity Range 10 to 90 Relative Humidity non condensing Setpoint Input Keypad Operation or External Setpoint Voltage Input Resolution Refer to Table 1 Setpoint Setp...

Page 8: ...lectrical connector Note 1 SLPM indicates the volumetric flow corrected to 20 C and 1 atmosphere 14 7 PSIA The reference temperature can also be changed to 0 C 25 C and 35 C The controllable flow rang...

Page 9: ...also displays the operation mode totalized flow rate value lower 4 digits valve drive output parameter setup values or function setup values OPERATION LAMP L Indicates the totalized flow rate is disp...

Page 10: ...nnection of a dedicated loader cable available separately to the loader jack on the rear of the MFC enables direct communication with a personal computer IN CONNECTOR 1 8 FNPT inlet connector OUT CONN...

Page 11: ...NG PROCEDURE PROCESS CONNECTION INSTALLATION While installing the inlet and outlet process connections to the inlet and outlet 1 8 FNPT connectors do not hold the housing of the MPC Series Panel Mount...

Page 12: ...From the front of the panel insert the MFC through the panel cutout until the MFC s bezel rests against the front of the panel With the mounting bracket screws oriented vertically i e 6 o clock and 12...

Page 13: ...ng power The event output does not exceed the specified output rating of the MFC When driving a relay use a relay with a built in diode for coil surge suppression ELECTRICAL CONNECTOR SPECIFICATIONS P...

Page 14: ...12 WIRING TYPICAL WIRING...

Page 15: ...e function setup For details on function setup method refer to SECTION 6 APPLICATION OPERATION 5 If no setting change is made while the multi setting flow rate is displayed the display will automatica...

Page 16: ...OW RATE DISPLAY When the DISP key is pressed while the operation mode is displayed the L lamp lights and the totalized flow rate value is indicated on the upper display and lower display For example i...

Page 17: ...when performing incremental changes to the SP value Multi SP Setpoint Setting Mode Multiple Setpoints Per Function Setup C 04 Up to four 4 SP setpoint values setting flow rate values can be programme...

Page 18: ...splayed indication is automatically returned to the instantaneous flow rate display PROCEDURE FOR CHANGING FLOW RATE IN ANALOG SETTING To change the SP value using an external setpoint voltage confirm...

Page 19: ...the DISP key is pressed 3 When the DISP key is pressed while the display is flashing step 2 of Selecting the Operating Mode procedure above the operating mode selection is cancelled 4 When 5 Valve ful...

Page 20: ...onds or longer press DISP key less than 1 second press DISP key or any key is not pressed for 10 seconds or longer press DISP key 1 second or longer flow rate indication 3 Totalized flow rate Totalize...

Page 21: ...ow rate display FUNCTION SETUP ITEM LIST Display Function Description Setup Options and Descriptions Factory Setting Remarks C 01 Key lock 0 Unlocked 1 Settings other than flow rate setting are locked...

Page 22: ...nal switch input 1 function assignment 0 C 11 External switch input 2 function assignment 0 Not used 1 Reset totalization 2 Stop totalization count operation 3 Switching of SP number 4 Switching of fl...

Page 23: ...pressure adjustment 0 0 0 1 MPa 0 14 5 PSIG 1 0 05 0 15 MPa 7 25 21 8 PSIG 2 0 15 0 25 MPa 21 8 36 3 PSIG 3 0 25 0 35 MPa 36 3 50 8 PSIG 4 0 35 0 45 MPa 50 8 65 3 PSIG 5 0 45 0 5 MPa 65 3 72 5 PSIG 2...

Page 24: ...ence No Display Parameter Description Factory Setting Setting Range Refer to Function Setup Item Remarks 1 0 r 9 1 Flow rate OK judgment range 2 FS 10 0 5 to 100 FS 10 2 0 HYS 1 Flow rate OK judgment...

Page 25: ...vent output type assignment in the function setup 7 This can be set only when 1 Function enabled is selected for the C 28 analog optional scaling function in the function setup 8 This can be set only...

Page 26: ...the gas is shut off for more than five 5 minutes set the set flow rate SP to zero or valve fully closed AL81 Sensor error Sensor problem particulate in sensor or use of helium or hydrogen gas Turn po...

Page 27: ...nstall an upstream filter Contact factory Flow rate does not stabilize Operational differential pressure range is exceeded Excessive supply pressure fluctuations Pressure regulator compatibility Exces...

Page 28: ...ter Instrument Division and accompanied by a properly completed Returned Goods Authorization All returns are subject to restocking and possible rework charges to be determined by Porter Instrument Div...

Page 29: ...ons and calibration the customer will be notified of the costs before proceeding with repair or replacement Repaired equipment will be warranted ninety 90 days from date of delivery to the customer or...

Page 30: ......