- 11 - 

TUBING CONNECTION 

 

 

When the MPC Series Panel Mount MFC is panel-mounted, use tubing that does not stress the MFC 
housing.  If metal piping is directly connected to the 1/8” FNPT connector of the MFC, the MFC 
should not be panel-mounted. 

 

Connect the tubing to ensure the gas flows in the direction from ‘IN’ to ‘OUT’ as indicated on the 
housing. 

 

After connecting the tubing, check for any gas leaks.  While performing the leak check, exercise 
caution to avoid spillage or contact of the leak detection fluid to the housing, electrical wiring or 
mating electrical connector.  If fluid contact occurs, MFC malfunction or faulty operation may occur. 

 

EXTERNAL ELECTRICAL CONNECTIONS 

 
Prior to wiring, confirm: 

 

Power supply source is off prior to wiring connection. 

 

Wiring is correct prior to introducing power. 

 

The event output does not exceed the specified output rating of the MFC.  When driving a relay, 
use a relay with a built-in diode for coil surge suppression. 

 

ELECTRICAL CONNECTOR SPECIFICATIONS 

 

 

Part No.:  MCVW1.5/9-STF-3.5 (Phoenix Contact Mfg.) 

 

Wire type: Either of single core wire or stranded wire applicable. 

 

Compatible cable: 0.08 to 1.5mm

2

 (SWF=28 to =16) 

 

Appropriate length of stripped wire: 7mm 

 

Compatible screw driver: Tip size 2.5 x 0.4mm (a flat-head driver) 

 

INPUT/OUTPUT DESIGNATIONS 

 

PIN 

NUMBER 

NAME/FUNCTION 

DESCRIPTION 

COMMENTS 

POWER (24V) 

Power (+24 Vdc) 

 

POWER GND 

Power supply ground 

 

EV1 

Event output 1 

EV2 

Event output 2 

Open collector non-isolated output 

DI1 

External switch input 1 

DI2 

External switch input 2 

Switching input (OPEN/GND) 

AI 

Analog setpoint voltage input 

A0 

Analog flow rate voltage 
output 

0 to 5 Vdc or 1 to 5 Vdc 
(selectable) 

SIGNAL GND 

Signal ground 

Input/output signal common. 
Signal ground is connected with 
power supply ground within MFC. 

 

Summary of Contents for Porter MPC Series

Page 1: ...MPC SERIES PANEL MOUNT MASS FLOW CONTROLLER INSTALLATION PROGRAMMING USER S MANUAL FM 1038 Rev 0 4 08...

Page 2: ...G 9 Location 9 Dimensional Data 9 Mounting Procedure 9 External Electrical Connections 11 Section 5 BASIC OPERATION 13 Switching Displays 13 Setting the Flow Rate 15 Selecting the Operation Mode 17 Se...

Page 3: ...positive shut off is a system requirement install a separate shut off valve When a separate shut off valve is closed the user must command the MFC control valve fully closed by one of the following me...

Page 4: ...User s Manual FM 1032 One 1 FEATURES The MPC Series Panel Mount MFC is a mass flow controller featuring high performance in a compact package design Integrating the advanced technologies of an ultra...

Page 5: ...rt stop reset can be remote controlled by external switching input After a reset by key operation the totalization calculation is started automatically However if a reset contact ON is carried out by...

Page 6: ...parate shut off valve AUTOMATIC RESET OF TOTALIZATION COUNT AT START OF CONTROL FUNCTION Start of control and totalization count reset can be carried out simultaneously by a single action key operatio...

Page 7: ...0 C 14 to 140 F Humidity Operating Humidity Range 10 to 90 Relative Humidity non condensing Setpoint Input Keypad Operation or External Setpoint Voltage Input Resolution Refer to Table 1 Setpoint Setp...

Page 8: ...lectrical connector Note 1 SLPM indicates the volumetric flow corrected to 20 C and 1 atmosphere 14 7 PSIA The reference temperature can also be changed to 0 C 25 C and 35 C The controllable flow rang...

Page 9: ...also displays the operation mode totalized flow rate value lower 4 digits valve drive output parameter setup values or function setup values OPERATION LAMP L Indicates the totalized flow rate is disp...

Page 10: ...nnection of a dedicated loader cable available separately to the loader jack on the rear of the MFC enables direct communication with a personal computer IN CONNECTOR 1 8 FNPT inlet connector OUT CONN...

Page 11: ...NG PROCEDURE PROCESS CONNECTION INSTALLATION While installing the inlet and outlet process connections to the inlet and outlet 1 8 FNPT connectors do not hold the housing of the MPC Series Panel Mount...

Page 12: ...From the front of the panel insert the MFC through the panel cutout until the MFC s bezel rests against the front of the panel With the mounting bracket screws oriented vertically i e 6 o clock and 12...

Page 13: ...ng power The event output does not exceed the specified output rating of the MFC When driving a relay use a relay with a built in diode for coil surge suppression ELECTRICAL CONNECTOR SPECIFICATIONS P...

Page 14: ...12 WIRING TYPICAL WIRING...

Page 15: ...e function setup For details on function setup method refer to SECTION 6 APPLICATION OPERATION 5 If no setting change is made while the multi setting flow rate is displayed the display will automatica...

Page 16: ...OW RATE DISPLAY When the DISP key is pressed while the operation mode is displayed the L lamp lights and the totalized flow rate value is indicated on the upper display and lower display For example i...

Page 17: ...when performing incremental changes to the SP value Multi SP Setpoint Setting Mode Multiple Setpoints Per Function Setup C 04 Up to four 4 SP setpoint values setting flow rate values can be programme...

Page 18: ...splayed indication is automatically returned to the instantaneous flow rate display PROCEDURE FOR CHANGING FLOW RATE IN ANALOG SETTING To change the SP value using an external setpoint voltage confirm...

Page 19: ...the DISP key is pressed 3 When the DISP key is pressed while the display is flashing step 2 of Selecting the Operating Mode procedure above the operating mode selection is cancelled 4 When 5 Valve ful...

Page 20: ...onds or longer press DISP key less than 1 second press DISP key or any key is not pressed for 10 seconds or longer press DISP key 1 second or longer flow rate indication 3 Totalized flow rate Totalize...

Page 21: ...ow rate display FUNCTION SETUP ITEM LIST Display Function Description Setup Options and Descriptions Factory Setting Remarks C 01 Key lock 0 Unlocked 1 Settings other than flow rate setting are locked...

Page 22: ...nal switch input 1 function assignment 0 C 11 External switch input 2 function assignment 0 Not used 1 Reset totalization 2 Stop totalization count operation 3 Switching of SP number 4 Switching of fl...

Page 23: ...pressure adjustment 0 0 0 1 MPa 0 14 5 PSIG 1 0 05 0 15 MPa 7 25 21 8 PSIG 2 0 15 0 25 MPa 21 8 36 3 PSIG 3 0 25 0 35 MPa 36 3 50 8 PSIG 4 0 35 0 45 MPa 50 8 65 3 PSIG 5 0 45 0 5 MPa 65 3 72 5 PSIG 2...

Page 24: ...ence No Display Parameter Description Factory Setting Setting Range Refer to Function Setup Item Remarks 1 0 r 9 1 Flow rate OK judgment range 2 FS 10 0 5 to 100 FS 10 2 0 HYS 1 Flow rate OK judgment...

Page 25: ...vent output type assignment in the function setup 7 This can be set only when 1 Function enabled is selected for the C 28 analog optional scaling function in the function setup 8 This can be set only...

Page 26: ...the gas is shut off for more than five 5 minutes set the set flow rate SP to zero or valve fully closed AL81 Sensor error Sensor problem particulate in sensor or use of helium or hydrogen gas Turn po...

Page 27: ...nstall an upstream filter Contact factory Flow rate does not stabilize Operational differential pressure range is exceeded Excessive supply pressure fluctuations Pressure regulator compatibility Exces...

Page 28: ...ter Instrument Division and accompanied by a properly completed Returned Goods Authorization All returns are subject to restocking and possible rework charges to be determined by Porter Instrument Div...

Page 29: ...ons and calibration the customer will be notified of the costs before proceeding with repair or replacement Repaired equipment will be warranted ninety 90 days from date of delivery to the customer or...

Page 30: ......

Reviews: