- 11 -
TUBING CONNECTION
When the MPC Series Panel Mount MFC is panel-mounted, use tubing that does not stress the MFC
housing. If metal piping is directly connected to the 1/8” FNPT connector of the MFC, the MFC
should not be panel-mounted.
Connect the tubing to ensure the gas flows in the direction from ‘IN’ to ‘OUT’ as indicated on the
housing.
After connecting the tubing, check for any gas leaks. While performing the leak check, exercise
caution to avoid spillage or contact of the leak detection fluid to the housing, electrical wiring or
mating electrical connector. If fluid contact occurs, MFC malfunction or faulty operation may occur.
EXTERNAL ELECTRICAL CONNECTIONS
Prior to wiring, confirm:
Power supply source is off prior to wiring connection.
Wiring is correct prior to introducing power.
The event output does not exceed the specified output rating of the MFC. When driving a relay,
use a relay with a built-in diode for coil surge suppression.
ELECTRICAL CONNECTOR SPECIFICATIONS
Part No.: MCVW1.5/9-STF-3.5 (Phoenix Contact Mfg.)
Wire type: Either of single core wire or stranded wire applicable.
Compatible cable: 0.08 to 1.5mm
2
(SWF=28 to =16)
Appropriate length of stripped wire: 7mm
Compatible screw driver: Tip size 2.5 x 0.4mm (a flat-head driver)
INPUT/OUTPUT DESIGNATIONS
PIN
NUMBER
NAME/FUNCTION
DESCRIPTION
COMMENTS
1
POWER (24V)
Power (+24 Vdc)
2
POWER GND
Power supply ground
3
EV1
Event output 1
4
EV2
Event output 2
Open collector non-isolated output
5
DI1
External switch input 1
6
DI2
External switch input 2
Switching input (OPEN/GND)
7
AI
Analog setpoint voltage input
8
A0
Analog flow rate voltage
output
0 to 5 Vdc or 1 to 5 Vdc
(selectable)
9
SIGNAL GND
Signal ground
Input/output signal common.
Signal ground is connected with
power supply ground within MFC.