Bulletin TI-NITROFLOW60-G
Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
9
1-800-343-4048
Figure 13
Figure 14
If the compressed air pressure decreases less than 90 psig (6.2 barg), the nitrogen output purity
level will decrease. Once pressure is restored, purity level will be restored.
The compressor output pressure gauge should read 115 psig to 120 psig during normal
operation.
See the Troubleshooting section of this manual on pages 17-20 to resolve any abnormal or faulty
conditions.
If the electrical power is interrupted, the generator will start automatically when the power
returns.
For normal or emergency shutdown, pull out the plug on the cordset to disconnect the mains
power (see Figure 1). Full depressurization takes 1 hour.
Do not remove cover while system
is pressurized.
Maintenance tasks for the NitroFlow60 and NitroFlow60D Generator should be performed by
trained personnel familiar with the service and safety precautions of electromechanical devices
to avoid injury or damage. Safety risks that may affect the service personnel are identified with
the necessary protective measures described.
Parker highly recommends service be performed by a Parker trained technician only. Call
the factory about PMP, the Preventative Maintenance Program. The service personnel will
verify the safe state of the generator after the maintenance or repair is completed.
Prior to servicing the generator, turn off the system and disconnect the cordset from the power
supply. Allow the system a minimum of 1 hour to depressurize. See shutdown section above.
To ensure product performance and reliability, periodic preventative maintenance must be
performed according to the following schedules (refer to the replacement parts section for
the part numbers when ordering); Schedule 1 is servicing the compressor and Schedule 2 is
servicing the generator.
Pressure Interruption
Power Interruption
Shutdown
Maintenance
The compressor has an hour meter to record the accumulated operating time of the compressor
for preventative maintenance scheduling of the serviceable components. The meter indicates
00000.0 hours. The hour meter is not re-settable.