Bulletin TI-NITROFLOW60-G
Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
15
1-800-343-4048
Carbon Modules
Particulate Filters
There are two carbon modules that require changing. Please note that the carbon modules will
only mount to the brackets in one orientation (see figure 28).
1
Remove the top panel from the NitroFlow60 Generator Series by removing two screws with a
Phillips head screwdriver.
2
Using a 1” wrench, remove the tie nut at the base of the bowl (see figure 30).
3
Twist the bowl loose from the head of the assembly and remove carefully as not to damage or
lose the o-ring (see figure 31).
4
Unscrew the element retainer at the base of the element, and replace the spent element with
a new one (see figure 32).
5
Reverse the steps to re-assemble. Make sure the o-ring is sealed in the groove before
reinstalling the bowl. Hand tighten.
Figure 28
Figure 30
Figure 31
Figure 32
Figure 29
1
Remove the top panel from the NitroFlow60 Generator Series by removing two screws with a
phillips head screwdriver and pulling top panel to the rear.
2
Disconnect the plastic tubing from the press fittings on the inlet and outlet ports of the module.
To remove the tubing from the press fitting, push the tubing into the fitting, hold the collar of
the fitting back, and pull the tubing out of the fitting.
3
Release the hook and loop strap that clamps the module to the unit (see figure 29). Slide the
module out of the bracket.
4
Replace with a new carbon module, noting the correct orientation (see figure 28).
5
Slide the new module into the bracket fully and re-attach the hook and loop strap.
6
Re-insert inlet and outlet tubing. Use a soap or a leak detecting solution to check for leaks
when pressurizing (re-starting) the unit.
Generator must be set to “OFF” and pressure gage must read zero before filter replacement.