Bulletin TI-NITROFLOW60-G
Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
16
1-800-343-4048
System Specifications
NitroFlow60
NitroFlow60D
CSA Certification Standard
CAN/CSA C22.2 No. 61010-1
CAN/CSA C22.2 No. 61010-1
IEC Standard
IEC 61010-1
IEC 61010-1
CENELEC Standard
EN 61010-1
EN 61010-1
UL Standard
UL 61010-1
UL 61010-1
IEC 61010 Installation
Category II
Category II
IEC 61010 Pollution
Degree 2
Degree 2
EMC Compliance
EN61326-1/CISPR 11
EN61326-1/CISPR 11
Nitrogen (N2) Purity
97%
97%
Nitrogen Max. Flow/Pressure
60 LPM/100 PSIG (6.9 Barg)
60 LPM/100 PSIG (6.9 Barg)
Dry Air Purity
-
-40° F (-40°C) ATM Dew Point
Dry Air Max. Flow/Pressure
-
4 SLPM/100PSIG (6.9 Barg)
Suspended Liquids
None
None
Particles > 0.01 Micron
None
None
Phthalates None
None
Sound Level
49 dB(A) at 1 Meter
49 dB(A) at 1 Meter
Min./Max. Operating Pressure
115/120 PSIG (7.9/8.2 Barg)
115/120 PSIG (7.9/8.2 Barg)
Flow Capacity
See Pressure/Flow Charts
See Pressure/Flow Charts
Inlet Air Consumption
160 SLPM (5.7 scfm )
160 SLPM (5.7 scfm)
Min./Max. Ambient Temp.
60°F/90°F (16°C-32°C)
60°F/90°F (16°C-32°C)
Max. Ambient Relative Humidity
80%
80%
Altitude 2000M
2000M
Heat Dissipation
10246 BTU/Hour
10246 BTU/Hour
Electrical Requirements¹ - 60 Hz
208-254VAC, 60 Hz, 1 Phase, 16A
208-254VAC, 60 Hz, 1 Phase, 16A
Electrical Requirements¹ - 50 Hz
230VAC, 50 Hz, 1 Phase, 13A
230VAC, 50 Hz, 1 Phase, 13A
Power Consumption - 60 Hz
16A @ 230VAC, 3680Watts
16A @ 230VAC, 3680Watts
Power Consumption - 50 Hz
13A @ 230VAC, 2990Watts
13A @ 230VAC, 2990Watts
Start-Up Time
10 Minutes
10 Minutes
Compressor Outlet - Tank Drain
1/4” Tube
1/4” Tube
Ports - Filter Drains (2 )
1/4” Tube
1/4” Tube
Ports - Water Seperator
1/4” Tube
1/4” Tube
Ports - Nitrogen Outlet
1/4” NPT
1/4” NPT
Ports - Dry Air Outlet
-
1/4” NPT
Fuses - Primary FU1, FU2
10.3 x 38.1mm, Type T, 600V, 1.25A
10.3 x 38.1mm, Type T, 600V, 1.25A
Fuses - Secondary FU3
6.3 x 31.7mm, Type T, 250V, 1.25A
6.3 x 31.7mm, Type T, 250V, 1.25A
Dimensions
33.5”L x 20.75”W x 45.25”H
33.5”L x 20.75”W x 45.25”H
Product Weight
407 lbs (185 kgs)
407 lbs (185 kgs)
Shipping Weight
632 lbs (287 kgs)
632 lbs (287 kgs)
¹ Main supply voltage fluctiations not to exceed ± 10% of nominal voltage.
2
Motor change installed in compressor Serial Number (H)8/8/2012-5284488-610 and after; NITROFLOW60 Serial Number, N260-0019 and after.
System Specifications
Cautions
1
The generator should be installed in an area with adequate ventilation to reduce the
flammability of the oxygen-rich permeate stream. The system should not be located in an
area where the permeate stream poses the risk of explosion or combustion
2
Nitrogen is nontoxic and largely inert. It can act as a simple asphyxiant by displacing
oxygen in air. Inhalation of nitrogen in excessive concentrations can result in
unconsciousness without any warning symptoms such as dizziness and fatigue
.
3
The maximum operating pressure of the system is 125 psig(8.6 barg). Operating the
generator at pressures above 125 psig(8.6 barg) will result in damage to the pump and
membrane.
4
The drain lines should be piped away to the provided collection container (open to
atmospheric pressure) to avoid any possible re-entrainment of liquid in the air which feeds
the generator.
5