Bulletin TI-NITROFLOW60-G
Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
7
1-800-343-4048
System
Comments Region
NitroFlow60NA
230V-60Hz, cordset has U.S. plug
North America
NitroFlow60DNA
230V-60Hz, cordset has U.S. plug
North America
NitroFlow60WD
230V-50Hz, cordset matched to region
Rest of world
NitroFlow60DWD
230V-50Hz, cordset matched to region
Rest of world
NitroFlow60JA200
230V-50Hz, cordset has a JA PSE plug
Japan
NitroFlow60DJA200
230V-50Hz, cordset has a JA PSE plug
Japan
The generator drain tube is a black nylon ¼” tubing connecting the drain outlet to the provided
disposal container (see Figure 8). A blue silencer is provided for the discharge end of the tubing
to reduce the discharge noise. Periodically check the built in strainer to ensure it has not clogged.
The pre-filter and water separator drain tubes are also a black nylon ¼” tubing connecting the
drain outlets to the same provided disposal container (see Figure 9). Blue silencers are provided
for each drain line.
The two outlet gas ports consist of two ¼” NPT bulkhead fittings (NitroFlow60D only. See Figure 10).
The non dry-air version consists of one ¼” NPT bulkhead fitting and a cap on the second outlet port.
Parker Balston recommends minimum 3/8” PFA plastic tubing with a minimum pressure rating of
125 psig (8.6 bar). Check all piping connections for leaks using a bubble-type leak detecting
solution.
Drains
Outlet Gases
Figure 8
There are six plastic tubing connections for the NitroFlow60 generator: (1) nitrogen outlet gas
connection; (1) dry air outlet gas connection (NitroFlow60D only); (4) drain connections. Press
the tubing into the quick-connect fittings. Assure tubing is fully inserted (about ½”). Tubing is
provided in the installation kit. For remote locations, additional length tubing is provided.
Piping
The following table describes the necessary cordset based on region.
Do not replace the provided
cordset with an inadequately rated cordset. The plug on the cordset is the mains disconnect.
The electrical schematic for the generator is shown in Figure 34 on page 22.
Notes
- “D” designation indicates additional dry air supply for chip cube requirements
- JA-200 versions include an external step-up transformer and gages indicating pressure in kPa
- WD(World Design) requires assignment of appropriate cordset, see table below
A03-0430
NA.............................................................30A, 250V
A03-0344
EU, IND, UK (ROW 230V,50Hz)...............32A, 250V
A03-0345
AU.............................................................32A, 250V
A03-0407
JA..............................................................30A, 250V
A03-0347
CHINA.......................................................32A, 250V
Cordset Rating
Figure 7, NEMA L6-30R 30AMP/250V
Receptacle for Model: NA (North America)