If necessary, wipe the generator with a clean, dry cloth on an as-needed basis. Do not use
water, aerosols or other cleaning agents on the unit. Use of any liquid detergent to clean
the unit could pose an electrical hazard.
Figure 18
Compressor Inlet Tube
Figure 19
Figure 20
Cleaning
Schedule 1 Instructions
1
Remove the upper side panel (see Figure 15).
2
By hand, push-in and rotate one intake filter cap to remove. Undo latch to remove second
intake filter cap (see Figure 20).
3
Replace the old filter elements with a new one, reverse steps to reassemble.
Intake Filter
1
Remove the top panel on the unit by removing two screws at the rear with a phillips head
screwdriver and pulling top panel to the rear.
2
Remove the generator cabinet on the unit by removing two screws with a phillips head
screwdriver.
a. LOOSEN (do not remove) three hex head screws located at the front of the cabinet
(figure 18).
b. Remove compressed air inlet tube from generator cabinet (see Figure 19).
c. Pull cabinet to the rear and lift off of compressor cabinet. Be cautious of exterior pre-filters.
3
Check belt tension (see figure 16) using a belt tension gauge or the deflection method.
Tension gauge should indicate minimum 45 lbs. If using the deflection method, 3.0-3.38 lbs
force at mid span should yield 7/32” belt deflection.
4
When making adjustments to the belts, make sure that pulley misalignment is no more than
1/16” and that both pulleys are not misaligned by more than 1/16” also (see Figure 21)
5
Adjust tension if required by adjusting motor bracket. Replace belts if needed.
Belts