6. Additional safety Information
Before operation
Ensure that the assembly is completed fully, and in accordance with the instructions given in the user manual.
Ensure that the local electrical mains supply voltage corresponds to that marked on the main control unit label
before connecting to the mains supply.
Ensure that cables from all motors are connected to the main control unit correctly before operation.
Ensure that all functions of the cot have been tested and that the bed is working normally.
IMPORTANT -
In the case of any unusual noises or smells during the operation,
immediately
switch off the power at
the mains and remove the mains plug from the power socket. Ensure that no cables are trapped in the mechanism or
are damaged. If you have any doubts concerning the safe operation of this cot, do not use the cot, and call your local
supplier or Park House Healthcare Ltd., Tel: 0845 0600 333
Persons assembling or using the cot should read the user manual and familiarise themselves with its operation.
Ensure that any objects such as waste paper baskets, occasional tables, chairs etc. are not placed beneath the
lifting mechanism of the cot.
To avoid injury, no limbs or body parts of either the occupant or carer must be protruding from the cot between the
side rails or into the mechanism of the bed during operation of powered functions.
Before moving the cot to a different location, remove from the mains power supply. Failure to do so may result in
damage to the bed and / or serious or fatal injury.
When moving the cot with occupant in-situ, do not push over threshold strips more than 2 cm max. height.
Under no circumstances must the cot be moved by pulling or pushing the side rails, either when the cot is occupied
or empty. Siderails are not designed for this purpose.
Never exceed the maximum usage period of 6 minutes per hour, for the electrically driven cot functions, by lengthy
and unnecessary profiling or height adjustments. This can trip the thermal safety fuse, which will stop powered
movement of all electrically operated cot parts.
Once the cot is sited in the desired position, always ensure that the castors are locked in
the braked position. (See photograph right).
Before dismantling or troubleshooting, ensure that all functions are in the lowered position
(where possible), the power supply is switched off and the power cable is disconnected
from the mains by un-plugging. Ensure that all functions are not under any undue load and
are free moving.
7.
Maintenance, care and warranty
To clean the powder coated tubular sections and wooden parts use a damp cloth with normal household cleaners or
warm soapy water.
Do not use any cleaning agents containing ammonia, abrasives or strong solvents. This will lead to damage of the
powder coated finish on metal parts and the finish on the wooden parts.
Mechanical cleaning, scouring, high-pressure hoses or automated cleaning will damage the cot.
Moving parts on the cot are fitted with self-lubricating bearings; they should not be oiled or greased.
All motors, cables and mechanical functions must be checked by a suitably qualified service engineer for damage
and correct operation, at least once every twelve months.