Park House Healthcare BUTTERFLY COT User Manual Download Page 10

10 

 

Extended warranties can be purchased from Park House Healthcare Ltd. For more information contact Customer 

Services on Tel: 0845 0600 333. In the event of a fault being discovered within the warranty period, the customer 

should  notify  the  Company  at  the  earliest  opportunity.  If,  upon  inspection,  the  Company  accepts  liability  then  the 

equipment shall be repaired or replaced 

at the company’s discretion. If the Company does not accept liability it will 

inform the customer of the reasons for rejection and provide the customer with an estimate of either the repair or 

replacement cost 

 

8.  Troubleshooting                                                                                                           

Fault 

Action 

Motors not working when 

handset operated. 

Check the connection to mains (when mains power supply is being 

used). Check the handset, motor and main control box connections. 

One motor does not work when 

the button is pressed. 

Check the connection between motor cable and control box. 

A BEEP sound is heard when 

the button is pressed. 

The 

‘backup battery’ (optional accessory) has low battery power.     

Plug power cable into mains socket and switch on. 

Handset or motors do not 

function. 

1) No power supply 

– Plug power cable into mains socket and turn on. 

2) Handset or motor defective - Contact Park House Healthcare Ltd. 

Motors suddenly have no 

response or movement. 

1) Switch off and unplug the mains cable; plug it back in and switch on. 

2) Leave the system in the rest position for approx. 20-30 min. 

3) Contact Park House Healthcare Ltd. 

Permanent or longer sounding warning signal. 

Contact Park House Healthcare Ltd. 

Handset LED lights up in locked state or does 

not light up in un-locked state. 

Contact Park House Healthcare Ltd. 

No adjustment can be made by 

handset. 

1) Try using the magnet key to release the locking function. 

2) Check the handset connections. 

3) Contact Park House Healthcare Ltd. 

WARNING! 

 

Any unauthorized modifications, adjustments or alterations will result in warranty invalidation.   

 

Repairs should only be carried out by a suitably qualified competent engineer. 

 

9.  Technical data / spare parts   

Overall length   

1800 mm 

Overall width 

860 mm 

Safe working load 

180 kg (28 Stone)   

Total Cot weight 

135kg / 21.3 Stones 

Sleeping platform height 

345mm~900mm   

  

  

Height of side rail (options) 

600mm / 900mm 

Castors 

12.5cm with central brake system 

Adjustment for backrest / Adjustment for leg rest 

66 deg +/- 2 deg        /      35 deg +/- 2 deg 

Power system 

Dewert actuator system 

Noise level 

Up to 65 dB(A) 

Raise-up motor 

Megamat MCZ 8000N, IP66 

Back motor 

Megamat MCZ 4000N, IP66 

Knee motor 

Megamat MCZ 4000N, IP66 

Control Box 

MCL3, IP66 

Handset 

IPROXX 10 button SE, IP66 

Backup battery   

AG7, IP66 

Tension voltage 

230v/50Hz 

Power duration 

Max. 6 min/h (10%) 

Summary of Contents for BUTTERFLY COT

Page 1: ...BUTTERFLY COT User Manual ...

Page 2: ...r sides 6 4 5 The door locking bar option 6 4 6 Connecting the cables 7 4 7 Correct positioning of the cables 7 5 Operating instructions 7 5 1 Operation of the height adjustment and profiling functions 7 5 2 The locking system of the handset 8 5 3 Using the lock out box functions 8 6 Additional Safety information 9 7 Maintenance care and warranty 9 8 Troubleshooting 10 9 Technical data spare parts...

Page 3: ...ot in the space between the undercarriage and mattress platform the adjustment functions on the handset must be locked to avoid unintentional activation of the mattress platform The mains power cable must be placed in a way so it does not get wound around the moving parts or trapped between them if the mains power cable is damaged there is a serious danger of injury from electrical current Inspect...

Page 4: ...ress Support 8 Side Rail 4 door bars 4 Head board Acrylic 9 Safety locking bar optional 4 1 Assembly of the sleeping platform The cot base lifting chassis and mattress support platform will be fully assembled upon delivery IMPORTANT Before operating the cot you must check that all cables have been fitted correctly and that they cannot become trapped damaged and dangerous during use Cot chassis wit...

Page 5: ...l The drop down side rail should be held in the closed position by ensuring that the side is fully raised and that the lower side lugs are located and seated securely in the holder A at each side at the bottom of each corner upright The side rail can be lowered by lifting the lower lugs out of the holders and carefully letting the rail slide down the side runners The side runners will move out sli...

Page 6: ...ting the securing latch upwards whilst opening the door For quick and easy full access to the side the doors can be folded and then fully swung back to each end of the cot 4 5 The door locking bar optional When using either type of the 4 door sides If greater security is required then the optional door locking bar can be installed and used There are pre drilled holes on the upper part of the insid...

Page 7: ...arts of the cot during use This could cause damage to the cables and create a serious safety hazard IMPORTANT Before using the cot and after re locating the lock out box check that all cables have been fitted correctly and that they can not become trapped damaged and dangerous during use 5 Operating instructions IMPORTANT Any person using this cot must read this instruction manual carefully and th...

Page 8: ...perator MUST ensure that Trendelenberg tilting will not have an adverse effect on the patient before using 5 3 Using the Lock out box functions WARNING When moving the lock out box location great care must be taken to ensure that the cable from the main control unit to the safety lock out box is positioned well clear of the lift mechanism and castors to eliminate the danger of shearing or crushing...

Page 9: ...serious or fatal injury When moving the cot with occupant in situ do not push over threshold strips more than 2 cm max height Under no circumstances must the cot be moved by pulling or pushing the side rails either when the cot is occupied or empty Siderails are not designed for this purpose Never exceed the maximum usage period of 6 minutes per hour for the electrically driven cot functions by le...

Page 10: ...lthcare Ltd Motors suddenly have no response or movement 1 Switch off and unplug the mains cable plug it back in and switch on 2 Leave the system in the rest position for approx 20 30 min 3 Contact Park House Healthcare Ltd Permanent or longer sounding warning signal Contact Park House Healthcare Ltd Handset LED lights up in locked state or does not light up in un locked state Contact Park House H...

Page 11: ... Protection The Butterfly Cot offers IP66 Ingress Protection on all electrical components and controls as standard This is a very high level of protection which is completely dustproof and capable of being washed by a jet of water without affecting the electrical operation of the bed WARNING The cot must be disconnected from the mains supply before any kind of washing procedure is undertaken This ...

Page 12: ...service and a network of distribution centres as part of our continued commitment to Customer Care 24 hour care 365 days a year dedicated internal personnel manage specific regions and divisions ensuring one on one personal customer service at all times With distributors in 25 over countries Park House Healthcare are a truly international company and we place important emphasis on providing high l...