4.4    The 4-door sides 

         

 

 

Each of the 4 doors has a securing latch in the bottom corner.   

 

The  doors  can  be  held  in  the  closed  position  by fully closing  them  and  ensuring  that  the  spring  latch is  securely 

located down into each of the holes located in the 4 door stop brackets, 

 

 

The doors can be further secured when closed by using the door lock which is fitted to each set of doors 

 

The  doors  can  be  easily  opened  by  firstly  ensuring  the  door  lock  is  unlocked  and  then  lifting  the  securing  latch 

upwards, whilst opening the door 

 

For quick and easy full access to the side, the doors can be folded and then fully swung back to each end of the cot.   

 

4.5      The door locking bar (optional) 

 

 

 

 

 

 

 

 

When using either type of the 4-door sides, If greater security is required then the optional door locking bar can be 

installed and used 

 

There  are  pre-drilled  holes  on the  upper  part  of  the inside  of  each  of the  4  doors,  which  can  be  used  to fix the  4 

locking bar brackets in the correct location. 

 

The locking bar can then be simply inserted down into the brackets, wrapping around each end of the cot. 

 

The bar can be quickly and easily removed by lifting it upwards 

Summary of Contents for BUTTERFLY COT

Page 1: ...BUTTERFLY COT User Manual ...

Page 2: ...r sides 6 4 5 The door locking bar option 6 4 6 Connecting the cables 7 4 7 Correct positioning of the cables 7 5 Operating instructions 7 5 1 Operation of the height adjustment and profiling functions 7 5 2 The locking system of the handset 8 5 3 Using the lock out box functions 8 6 Additional Safety information 9 7 Maintenance care and warranty 9 8 Troubleshooting 10 9 Technical data spare parts...

Page 3: ...ot in the space between the undercarriage and mattress platform the adjustment functions on the handset must be locked to avoid unintentional activation of the mattress platform The mains power cable must be placed in a way so it does not get wound around the moving parts or trapped between them if the mains power cable is damaged there is a serious danger of injury from electrical current Inspect...

Page 4: ...ress Support 8 Side Rail 4 door bars 4 Head board Acrylic 9 Safety locking bar optional 4 1 Assembly of the sleeping platform The cot base lifting chassis and mattress support platform will be fully assembled upon delivery IMPORTANT Before operating the cot you must check that all cables have been fitted correctly and that they cannot become trapped damaged and dangerous during use Cot chassis wit...

Page 5: ...l The drop down side rail should be held in the closed position by ensuring that the side is fully raised and that the lower side lugs are located and seated securely in the holder A at each side at the bottom of each corner upright The side rail can be lowered by lifting the lower lugs out of the holders and carefully letting the rail slide down the side runners The side runners will move out sli...

Page 6: ...ting the securing latch upwards whilst opening the door For quick and easy full access to the side the doors can be folded and then fully swung back to each end of the cot 4 5 The door locking bar optional When using either type of the 4 door sides If greater security is required then the optional door locking bar can be installed and used There are pre drilled holes on the upper part of the insid...

Page 7: ...arts of the cot during use This could cause damage to the cables and create a serious safety hazard IMPORTANT Before using the cot and after re locating the lock out box check that all cables have been fitted correctly and that they can not become trapped damaged and dangerous during use 5 Operating instructions IMPORTANT Any person using this cot must read this instruction manual carefully and th...

Page 8: ...perator MUST ensure that Trendelenberg tilting will not have an adverse effect on the patient before using 5 3 Using the Lock out box functions WARNING When moving the lock out box location great care must be taken to ensure that the cable from the main control unit to the safety lock out box is positioned well clear of the lift mechanism and castors to eliminate the danger of shearing or crushing...

Page 9: ...serious or fatal injury When moving the cot with occupant in situ do not push over threshold strips more than 2 cm max height Under no circumstances must the cot be moved by pulling or pushing the side rails either when the cot is occupied or empty Siderails are not designed for this purpose Never exceed the maximum usage period of 6 minutes per hour for the electrically driven cot functions by le...

Page 10: ...lthcare Ltd Motors suddenly have no response or movement 1 Switch off and unplug the mains cable plug it back in and switch on 2 Leave the system in the rest position for approx 20 30 min 3 Contact Park House Healthcare Ltd Permanent or longer sounding warning signal Contact Park House Healthcare Ltd Handset LED lights up in locked state or does not light up in un locked state Contact Park House H...

Page 11: ... Protection The Butterfly Cot offers IP66 Ingress Protection on all electrical components and controls as standard This is a very high level of protection which is completely dustproof and capable of being washed by a jet of water without affecting the electrical operation of the bed WARNING The cot must be disconnected from the mains supply before any kind of washing procedure is undertaken This ...

Page 12: ...service and a network of distribution centres as part of our continued commitment to Customer Care 24 hour care 365 days a year dedicated internal personnel manage specific regions and divisions ensuring one on one personal customer service at all times With distributors in 25 over countries Park House Healthcare are a truly international company and we place important emphasis on providing high l...