4.6    Connecting the cables                                                                             

 

The 

Control box

 has colour coding to clearly indicate the correct cable connections. 

IMPORTANT 

  Please  ensure  that  all  cables  are  connected  correctly, and  test  all of  the functions  fully 

before using the cot 

 

Only connect the components as shown.   

Connecting differently may result in damage to the control unit. 

Colour Codes 

1. 

Green: Lock-out device. 

2.  Red

: Handset 

3.  Black

: Backrest motor. 

4.  Yellow

: Footrest motor. 

5.  Blue:

 Head raise-up motor 

6.  White

: Foot raise-up motor 

7.  Black

: Backup battery plug. 

 

4.7    Correct positioning of the cables                                                                             

 

 

The cables must be connected in the correct manner to avoid trapping the cables between the metal parts of the cot 

during use. This could cause damage to the cables and create a serious safety hazard. 

 

IMPORTANT 

– 

Before  using  the  cot,  and  after  re-locating the  lock-out  box,  check that  all  cables  have  been fitted 

correctly and that they can not become trapped, damaged and dangerous during use.

 

 

 

5. 

Operating instructions       

                                                                                   

IMPORTANT

  -  Any  person  using  this  cot  must  read  this  instruction  manual  carefully  and  they  should  be  fully 

conversant with the correct operating procedures for its safe operation. 

 

5.1 Operation of the height adjustment and profiling functions                                                       

The head section, knee break, cot height and Trendelenburg tilting are adjusted by pressing the appropriate button on 

the handset. For ease of use, the buttons are labelled graphically according to their function. 

 

WARNING

 - Do not exceed the maximum usage period 

– all functions maximum of 6 minutes per hour.

 

 

 

 

 

 

 

 

Summary of Contents for BUTTERFLY COT

Page 1: ...BUTTERFLY COT User Manual ...

Page 2: ...r sides 6 4 5 The door locking bar option 6 4 6 Connecting the cables 7 4 7 Correct positioning of the cables 7 5 Operating instructions 7 5 1 Operation of the height adjustment and profiling functions 7 5 2 The locking system of the handset 8 5 3 Using the lock out box functions 8 6 Additional Safety information 9 7 Maintenance care and warranty 9 8 Troubleshooting 10 9 Technical data spare parts...

Page 3: ...ot in the space between the undercarriage and mattress platform the adjustment functions on the handset must be locked to avoid unintentional activation of the mattress platform The mains power cable must be placed in a way so it does not get wound around the moving parts or trapped between them if the mains power cable is damaged there is a serious danger of injury from electrical current Inspect...

Page 4: ...ress Support 8 Side Rail 4 door bars 4 Head board Acrylic 9 Safety locking bar optional 4 1 Assembly of the sleeping platform The cot base lifting chassis and mattress support platform will be fully assembled upon delivery IMPORTANT Before operating the cot you must check that all cables have been fitted correctly and that they cannot become trapped damaged and dangerous during use Cot chassis wit...

Page 5: ...l The drop down side rail should be held in the closed position by ensuring that the side is fully raised and that the lower side lugs are located and seated securely in the holder A at each side at the bottom of each corner upright The side rail can be lowered by lifting the lower lugs out of the holders and carefully letting the rail slide down the side runners The side runners will move out sli...

Page 6: ...ting the securing latch upwards whilst opening the door For quick and easy full access to the side the doors can be folded and then fully swung back to each end of the cot 4 5 The door locking bar optional When using either type of the 4 door sides If greater security is required then the optional door locking bar can be installed and used There are pre drilled holes on the upper part of the insid...

Page 7: ...arts of the cot during use This could cause damage to the cables and create a serious safety hazard IMPORTANT Before using the cot and after re locating the lock out box check that all cables have been fitted correctly and that they can not become trapped damaged and dangerous during use 5 Operating instructions IMPORTANT Any person using this cot must read this instruction manual carefully and th...

Page 8: ...perator MUST ensure that Trendelenberg tilting will not have an adverse effect on the patient before using 5 3 Using the Lock out box functions WARNING When moving the lock out box location great care must be taken to ensure that the cable from the main control unit to the safety lock out box is positioned well clear of the lift mechanism and castors to eliminate the danger of shearing or crushing...

Page 9: ...serious or fatal injury When moving the cot with occupant in situ do not push over threshold strips more than 2 cm max height Under no circumstances must the cot be moved by pulling or pushing the side rails either when the cot is occupied or empty Siderails are not designed for this purpose Never exceed the maximum usage period of 6 minutes per hour for the electrically driven cot functions by le...

Page 10: ...lthcare Ltd Motors suddenly have no response or movement 1 Switch off and unplug the mains cable plug it back in and switch on 2 Leave the system in the rest position for approx 20 30 min 3 Contact Park House Healthcare Ltd Permanent or longer sounding warning signal Contact Park House Healthcare Ltd Handset LED lights up in locked state or does not light up in un locked state Contact Park House H...

Page 11: ... Protection The Butterfly Cot offers IP66 Ingress Protection on all electrical components and controls as standard This is a very high level of protection which is completely dustproof and capable of being washed by a jet of water without affecting the electrical operation of the bed WARNING The cot must be disconnected from the mains supply before any kind of washing procedure is undertaken This ...

Page 12: ...service and a network of distribution centres as part of our continued commitment to Customer Care 24 hour care 365 days a year dedicated internal personnel manage specific regions and divisions ensuring one on one personal customer service at all times With distributors in 25 over countries Park House Healthcare are a truly international company and we place important emphasis on providing high l...

Reviews: