
Gener
al
Information
Pr
epar
e
Oper
ate
Help
Technology
Electromagnetic compatibility (EMC)
151
Manufacturer's declaration – electromagnetic immunity
The PR 25 is designed to be operated under the electromagnetic conditions specified below. The customer
or the user of the device should ensure that it is used under these conditions.
Interference immunity
tests
IEC 60601 – test level
Compliance level
Electromagnetic conditions –
guidelines
Electrostatic discharge
(ESD), in accordance
with IEC 61000-4-2
± 6 kV Contact
discharge
± 8 kV Air discharge
± 6 kV Contact
discharge
± 8 kV Air discharge
Floors should be made from wood
or concrete, or have ceramic tiles. If
the flooring is made from synthetic
material, the relative humidity must
be at least 30%.
Fast transient electric
bursts in accordance
with IEC 61000-4-4
± 2 kV for power supply
cables
± 1 kV for input and
output cables
± 1 kV for input and
output cables
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
Surges
in accordance with
IEC 6100-4-5
± 1 kV differential mode
voltage
± 2 kV common-mode
voltage
Not usable
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
Summary of Contents for WHEELCHAIR Series
Page 1: ...Operating instructions PR 25 PARAVAN WHEELCHAIR SERIES EN V1 40...
Page 158: ......
Page 159: ......