Gener
al
Information
Pr
epar
e
Oper
ate
Help
Technology
Loading and transporting the power wheelchair
93
13.5.7 Manual release in case of electric failure
A manual emergency release is located at the front edge of the
docking station.
>
Move wheelchair forward to remove the pressure on the lock pin
and push the red release arm to one side and hold it there while the
wheelchair moves away.
>
A cable-activated manual operating lever can also be fitted
(accessory). The red release arm is also pushed to one side and
should be held there whilst the wheelchair moves away.
If the described manual release procedures fails, an emergency
release tool made from red plastic comes with each docking station.
>
Move wheelchair forward to remove the pressure on the lock pin
>
Place the emergency release tool in the gap between the locking
plate and the docking station.
>
Push the release tool and wheelchair forward until the locking pin
has been forced down - after which the wheelchair can reverse out
of the docking station.
>
Please also refer to Dahl Engineering instructions for installa-
tion, use and maintenance for Dahl Docking Station MK II and
VarioDock™ systems.
Fig. 46: Manual release step 1
Fig. 47: Manual release step 2
Summary of Contents for PR 35
Page 1: ...User manual PR 35 PARAVAN WHEELCHAIR SERIES EN V1 23...
Page 160: ...General Information Prepare Operate Help Technology Annexes and technical documentation 160...
Page 161: ...General Information Prepare Operate Help Technology Annexes and technical documentation 161...
Page 162: ...General Information Prepare Operate Help Technology Annexes and technical documentation 162...
Page 163: ......