Manufacturer's declaration – electromagnetic immunity
The PR 35 is designed to be operated under the electromagnetic conditions specified below. The customer
or the user of the device should ensure that it is used under these conditions.
Interference immunity
tests
IEC 60601 – test level
Compliance level
Electromagnetic conditions – guide-
lines
Electrostatic discharge
(ESD), in accordance
with IEC 61000-4-2
± 6 kV Contact dischar-
ge
± 8 kV Air discharge
± 6 kV Contact dischar-
ge
± 8 kV Air discharge
Floors should be made from wood
or concrete, or have ceramic tiles. If
the flooring is made from synthetic
material, the relative humidity must
be at least 30%.
Fast transient electric
bursts in accordance
with IEC 61000-4-4
± 2 kV for power supply
cables
± 1 kV for input and
output cables
± 1 kV for input and
output cables
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
Surges, in accordance
with IEC 6100-4-5
± 1 kV differential mode
voltage
± 2 kV common-mode
voltage
Not usable
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
Summary of Contents for PR 35
Page 1: ...User manual PR 35 PARAVAN WHEELCHAIR SERIES EN V1 23...
Page 160: ...General Information Prepare Operate Help Technology Annexes and technical documentation 160...
Page 161: ...General Information Prepare Operate Help Technology Annexes and technical documentation 161...
Page 162: ...General Information Prepare Operate Help Technology Annexes and technical documentation 162...
Page 163: ......