7
Maintenance
98
6924-en_V1.3_09/20
Entering flow rates
95,00
X,00
salt | oil
Menu
test bench
1.
Ensure that the filter holder is closed.
2.
Open the menu
test bench
.
3.
Verify that the correct medium for the generator is entered. “
salt
” for
saline aerosol / “
oil”
for oil aerosol
4.
Enter the value
95
l/min for the parameter
main air flow sp
. This
value is the flow rate Q that is used in the formula.
5.
Enter the standard value for the parameter
generator flow sp
:
7
l/min for saline aerosol /
2.5
l/min for oil aerosol