4
Initial commissioning
6924-en_V1.3_09/20
25
Connections on
VKL 100
and
Promo 1000
8
2
5
7
6
3
4
1
Connections on VKL 100 and Promo 1000
1
Promo 1000
2 Hose line (NW 6 mm) from out-
let of aerosol sensor
we-
las 1200
to plug connection
“Suction”
3 Line “Detector” to aerosol sen-
sor
4 Line “Lamp” to aerosol sensor
5 Compressed air line
(NW 13 mm), 6 – 8 bar, from
the back of the test rig
6 Exhaust fitting for hose ID
28 mm
7 Dilution system
VKL 100
8 Sampling tube
The exhaust fitting must be connected when depressurized to a filter and pu-
rification system belonging to the operator (system not included in delivery).
WARNING
Escaping aerosols might be toxic
Aerosols can be hazardous to human health. If the aerosol outlet is not
connected to a suitable filter or purification system, aerosols can escape
into the environment and harm persons.
4
Before switching on: Verify that all aerosol outlets are connected to a
suitable filter and purification system.