Testing procedure
There are two ways to place respiratory masks in the filter holder:
• With the exterior (curvature) facing up: simulates inhalation.
• With the exterior (curvature) facing down: simulates exhalation.
The software contains the following testing routines:
• Efficiency testing with a saline aerosol
• Efficiency testing with an oil aerosol
• Breath resistance testing without aerosol
• Testing of valve on FFP masks without aerosol
Unusually shaped respiratory masks can be placed on the included universal
filter adapter and tested there.
Flat filter media can be placed on an included support grid and tested there.
3.6
Environmental conditions
Environmental conditions affect the functionality of the device.
Operation
The following environmental conditions apply for the operation of the device:
• Temperature range: +19°C to +23°C
• Relative humidity: non-condensing, meaning that moisture in the air is
not permitted to condense as water on or in the device
• Air pressure: atmospheric pressure
Storage and transport
The following environmental conditions apply for the storage and transport of
the device:
• Temperature range: 0°C to +40°C
Operation and storage under other environmental conditions, for instance in
corrosive or explosive environments, powerful electrical or electromagnetic
fields, areas with ionizing radiation or areas subject to shocks and vibrations
are prohibited.