7
Maintenance
6924-en_V1.3_09/20
85
7.2.1
Preparations for calibration
To calibrate the aerosol sensor, use the firmware
Mas-Q-Check
of the
aerosol spectrometer (
Promo 1000
or
Promo LED 2300
). The firmware can
be operated either directly on the touch screen of the aerosol spectrometer
or via the remote connection.
The software
PMFT1000
has to be closed to be able to use a remote con-
nection. The two applications cannot be used at the same time.
Opening the menu
settings
in the firmware
Mas-Q-Check
of
the aerosol spectrometer
4
Open the menu
settings
in the firmware
Mas-Q-Check
.
ð
The menu
settings
opens.
Mas-Q-Check firmware: Menu
settings