20
Table 2 compliance standards
Phenomenon
Basic
Standard of
Immunity Test
Level of
Electrostatic
discharge
IEC 61000-4-2
±8 KV contact
±2 KV, ±4 KV,
±8 KV,±15KV air§
±8 KV contact
± 2 K V, ± 4 K V,
± 8 KV, ±15KV air
Radiated RF EM
fields1
IEC 61000-4-3
3V/m
80 MHz-2.7 GHz
80% AM at 1 KHz
3V/m
80 MHz-2.7 GHz
80% AM at 1 KHz
Proximity
fields from
RF wireless
communication
equipment
IEC 61000-4-3
See table
See table
Electrical Fast/
Transients
bursts
IEC 61000-4-4
±1 KV
100 KHz
repetition
frequency
±1 KV
100 KHz
repetition
frequency
Conducted
disturbances
induced by RF
fields.
IEC 61000-4-6
3V
0.15 MHz-80
MHz 6 Vm in ISM
bands between
0 .15 MHz and
80 MHz
80% AM at 1KHz
3V
0.15 MHz-80
MHz 6 Vm in ISM
bands between
0.15 MHz and 80
MHz
80% AM at 1KHz
Rated power
frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz