SOURCES OF MEASUREMENT ERRORS
When making flux density measurements there are several conditions that can introduce error:
1) Operating the meter while the LOW BATTERY symbol appears.
Instrument specifications are not guaranteed when a low battery condition exists!
2) Failure to zero the error signals from the meter, probe, and nearby sources of magnetic interference.
3) Subjecting the probe to physical abuse.
4) One of the most common sources of error is the angular position of the probe with respect to the field being measured.
As mentioned in Section1, a Hall generator is not only sensitive to the number of flux lines passing through it but also to the
angle which the pass through it. The Hall generator produces the greatest signal when the flux lines are perpendicular to
the sensor as shown in Figure 3-22.
The probe is calibrated and specified with flux lines passing perpendicularly through the Hall generator.
5) As shown in Figure 3-23 the greater the distance between the magnetic source and the Hall probe, the fewer flux lines
will pass through the probe. This would cause the probe's output to decrease.
Handle the Hall probe with care. Do not bend the stem or apply pressure to the probe tip as
damage may result. Use the protective cover when the probe is not in use.