Figure 3-18
MANUAL RELATIVE Function
Handle the Hall probe with care. Do not bend the stem or apply
pressure to the probe tip as damage may result.
MANUAL RELATIVE MODE
The second method by which to set a relative value is a manual adjustment. In some cases the user will wish to set an absolute
relative value. To do this, insert the probe in the "zero flux chamber" provided with this meter.
To manually assign a relative offset value, press the SHIFT key followed by the REL key. At this point the "RELATIVE" icon will be
flashing and actual flux density readings will be displayed. Use the LEFT, RIGHT, UP, and DOWN arrow keys to select a desired
offset. When complete press the SHIFT key followed by the REL key to exit and return to the measurement mode. The RELATIVE
icon will be on indicating that a relative offset is being subtracted from all incoming readings.
In other cases, the probe may be positioned in a stable DC or AC magnetic field that does not exceed the present range limit. In
most cases the reading will be adjusted to zero so that the value of the reference is subtracted from all incoming readings.
To deactivate the manual relative mode, press the SHIFT key followed by the REL key. The meter will enter the MANUAL RELATIVE
mode. Press the LEFT arrow key until the OFF icon turns on, then press the SHIFT key followed by the REL key to exit and return to
the measurement mode. Alternatively, from within the MANUAL RELATIVE mode, Press the RESET key to clear the relative offset
value. Press the SHIFT key followed by the REL key to exit and return to the measurement mode.