Figure 3-13
HOLD Function
HOLD MODE SELECTION
In some applications it may be desirable to hold a reading that is either greater than or less than all previous readings, or
which has the greatest peak value whether positive or negative.
The MAX HOLD function holds the reading that is arithmetically greater than all previous readings. For instance a reading of
+125.0 is greater than +99.0 or -150.0.
The MIN HOLD function holds the reading that is arithmetically less than all previous readings. For instance a reading of
-125.0 is less than 99.0 or +150.0.
The PEAK HOLD function captures and holds the peak value of the flux density waveform within the response time
capabilities of the meter. See the
SPECIFICATIONS
section of this manual for more information. The peak can be either
positive or negative, whichever has the greatest magnitude. For instance a peak value of -100.0 is greater than a peak
value of +90.0.
To choose the desired hold mode press the SHIFT pushbutton followed by the RESET pushbutton. Press the LEFT (3) and
RIGHT (4) arrows to select the desired mode on the display. MAX HOLD is indicated by “MX”. MIN HOLD is indicated by
“MN”. PEAK HOLD is indicated by “PK”. The “OFF” legend will appear when all HOLD modes are turned off.
Difference between PEAK and MAX for AC: Max is the maximum rms value. PEAK gives the value of the point on the
waveform with the greatest amplitude.
3-8