Figure 3-11
UNITS Function
Figure 3-8
Low Battery Indication
Figure 3-9
Overrange Indication
Figure 3-10
MODE (AC/DC) Function
LOW BATTERY CONDITION
The meter is designed to use four standard AA 1.5V alkaline batteries . When the battery
voltage becomes too low the battery symbol on the display will flash, as shown in Figure 3-8.
Replace the batteries or use an external AC to DC power supply.
Instrument specifications are not guaranteed when a low battery condition exists!
OVERRANGE CONDITION
If the magnitude of the magnetic flux density exceeds the limit of the selected range the meter
will display a flashing value of “2999” (gauss or tesla mode) or “2387” (ampere meter mode).
The next highest range should be selected. If already on the highest range then the flux
density is too great to be measured with this instrument.
AC OR DC MEASUREMENT SELECTION
The meter is capable of measuring either static (DC) or alternating (AC) magnetic fields. To
choose the desired mode Press the SHIFT pushbutton followed by the right (4) arrow to select
AC or DC on the display. The DC and AC mode is discussed in more detail later in this
section. This setting is saved and will be restored the next time the meter is turned on.
UNITS OF MEASUREMENT SELECTION
The meter is capable of providing flux density measurements in terms of gauss (G), tesla (T)
or ampere/meters (A/m). To choose the desired units, Press the SHIFT pushbutton followed
by the left (3) arrow pushbutton. Press the up (
5
) and down (
6
) arrow pushbuttons to select G,
T or Am on the display. This setting is saved and will be restored the next time the meter is
turned on.
3-6