SECTION 5
REMOTE OPERATION
78
5.5.2 METERED DATA QUERIES
Measurement occurs once per :MEASure: command. AC measurements are assumed.
The SCPI
:AC
keyword following
:MEASure
is optional and is not shown here.
The “
?
” is optional as
:MEASure:
implies a query. Values following a command are ignored.
'Multiple keyword' queries may be performed to obtain values of simultaneous measurements. e.g.,
:MEASure:VOLTage1;VOLTage2;VOLTage3?
:MEASure:VOLTage1?
Query: metered voltage
Ф
A
Returns <0> to <354> VAC RMS
:MEASure:VOLTage2?
Query: metered voltage
Ф
B
Returns <0> to <354> VAC RMS
:MEASure:VOLTage3?
Query: metered voltage
Ф
C
Returns <0> to <354> VAC RMS
:MEASure:VLL1?
Query: metered Line to Line voltage
Ф
A to
Ф
B
Returns <0> to <354> VAC RMS
:MEASure:VLL2?
Query: metered Line to Line voltage
Ф
B to
Ф
C
Returns <0> to <354> VAC RMS
:MEASure:VLL3?
Query: metered Line to Line voltage
Ф
C to
Ф
A
Returns <0> to <354> VAC RMS
:MEASure:FREQuency?
Query: for programmed Frequency value
Returns <20.00> to <5000> Hz
The UPC reports only the programmed value, it does not measure
frequency
:MEASure:CURRent[:RMS]1?
Query: metered RMS current
Ф
A
Returns <0> to <current meter range> AAC RMS
:MEASure:CURRent[:RMS]2?
Query: metered RMS current
Ф
B
Returns <0> to <current meter range> AAC RMS
:MEASure:CURRent[:RMS]3?
Query: metered RMS current
Ф
C
Returns <0> to <current meter range> AAC RMS
:MEASure:CURRent:PEAK1?
Query: metered peak current
Ф
A
Returns <0> to <current meter range> A Peak
:MEASure:CURRent:PEAK2?
Query: metered peak current
Ф
B
Returns <0> to <current meter range> A Peak
:MEASure:CURRent:PEAK3?
Query: metered peak current
Ф
C
Returns <0> to <current meter range> A Peak