SECTION 4
OPERATION
51
4.5.2.2 CYCLE BASED TRANSIENT EXAMPLE (continued)
Press
STORE,
to save the transient sequence.
NUMBER OF TRANSIENT EVENTS=1
(0 for infinite or 1-65535)
TRANSIENT AUTO RMS=ON
(Press +/- Key to Toggle Selection)
Press
1, ENTER,
to cause only one transient event.
NUMBER OF TRANSIENT EVENTS=1
(0 for infinite or 1-65535)
TRANSIENT AUTO RMS=ON
(Press +/- Key to Toggle Selection)
Press
STORE,
to save Transient with AUTO RMS active or
+/-
,
STORE
to disable AUTO RMS.
The display now shows V/I METER.
V/I METER:
FREQ=400.0 Va=108.0 Vb=108.0 Vc=108.0
SENSE=INT Vab=187.1 Vbc=187.1 Vca=187.1
MANUAL MODE Ia=00.0 Ib=00.0 Ic=00.0
If the UPC was previously executing program #4, it will now be in MANUAL MODE, as indicated at the lower
left corner of the display, because program #4 has changed since it was EXECUTED. To use the newly
created transient you must first EXECUTE Program #4 again. Press
PROGRAM, 4, ENTER, EXECUTE,
to
execute the program.
V/I METER:
FREQ=400.0 Va=108.0 Vb=108.0 Vc=108.0
SENSE=INT Vab=187.1 Vbc=187.1 Vca=187.1
PROGRAM #4 Ia=00.0 Ib=00.0 Ic=00.0
Press
TRANS
to execute the transient.
This cycle based transient will switch waveform #15 in for one cycle. This will cause a spike on the positive
and negative half of the sine wave for one cycle (
).