SECTION 4
OPERATION
43
4.5.1.1 PROGRAM EXAMPLE (continued)
Press
108, ENTER, 108, ENTER, 108, ENTER,
to select 108 volts line to neutral on all three phases.
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=400 Va=108 Vb=108 Vc=108
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120° PHc=240°
Press
1, ENTER, 1, ENTER, 1, ENTER,
to select WaveForm 1 on all three phases.
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=400 Va=108 Vb=108 Vc=108
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120° PHc=240°
Press
ENTER, 50, ENTER
to set Ilim
If Current Protect has been activated (
), then, after setting Ilim:
Press
CLEAR, CLEAR, +/-
key to show IprA, press
ENTER,
value,
ENTER
to set IprA.
Press
CLEAR, CLEAR, +/-
key to show IprT, press
ENTER,
value,
ENTER
to set IprT
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=400 Va=108 Vb=108 Vc=108
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120° PHc=240°
Press
120, ENTER, 240, ENTER,
to select 120° phase shift for phase B and 240° phase shift for phase C
referenced to phase A.
PROGRAM: #4 FORM=3 COUPLING=DIRECT
FREQ=400 Va=108 Vb=108 Vc=108
WFa=1 WFb=1 WFc=1
#SEGS=0 Ilim=50 PHb=120° PHc=240
Press
EXECUTE
to execute this new program. The V/I METER display will become active.
V/I METER:
FREQ=400.0 Va=108.0 Vb=108.0 Vc=108.0
SENSE=INT Vab=187.1 Vbc=187.1 Vca=187.1
PROGRAM #4 Ia=00.0 Ib=00.0 Ic=00.0