![Oxford Instruments X-MET8000 Series Operator'S Manual Download Page 26](http://html1.mh-extra.com/html/oxford-instruments/x-met8000-series/x-met8000-series_operators-manual_1664946026.webp)
Alloy+ application, Fundamental Parameter For Gold
How Long For A Measurement?
The factory set measurement time of 10 seconds is applicable for many measurements. Quicker
measurements for iron, copper, nickel and similar alloys require as little as 5 seconds. Magnesium
and aluminum alloys need longer measurement times, for example 15 seconds or more. Complex
alloys also require longer measurement times to analyze all the trace elements.
Longer measurements will always give more precise results. However, the X-MET8000 series can
provide excellent results in only a few seconds. The standard deviation, shown as +/- on the display,
figures indicate the measurement precision. The longer the measurement time, the lower the standard
deviation figures, the greater the precision.
26
X-MET8000 Series Operator’s Manual
8