HDS2062M-N Handheld DSO & DMM 8-Trouble Shooting
z
Try to change the trigger coupling into the HF rejection and LF rejection to filter the HF
or LF noise trigger by the interference.
6.
There is no display on the screen when you push down
RUN/STOP key under the oscilloscope mode.
Check whether the trigger mode in the trigger mode menu is in normal or single shot and
whether the trigger level is out of the waveform range.
In such condition, adjust and make the trigger level in the middle or select the non-auto
trigger mode. Additionally, press the AUTO SET key and complete the above setting
7.
When select the average sampling in the sampling mode
or select a longer display time in the display mode under
the oscilloscope, the display speed is slow.
It is normal in the above case.
61