HDS2062M-N Handheld DSO & DMM
7-Advanced Function of Oscilloscope
displayed on the screen in green color.
The display color is green, and the zero point of waveform, voltage and time is purple
Now, you can see a screen that looks like the following figure 45.
Figure 45: Waveform Saving
7.9 Function Setting Menu
The
function setting
menu is described in the following list.
Function menu Setting
Description
Recall Factory
Resume the instrument to its factory settings.
Auto
Calibration
Perform the Auto-calibration procedure.
Language Chinese
English
Select the display language.
Auto-calibration
The Auto-calibration function automatically configures internal parameters to maintain the
sensitivity and accuracy. Run the Auto-calibration in the following cases:
When the temperature fluctuates more than 5 degrees Celsius during operations.
When operating the oscilloscope in a new bench top or field environment.
Procedure:
1. Press the MENU key and select the
FUNCTION
menu using
MENU
▲
or
MENU
▼
key.
2. Press F2 (Auto calibration). A message appears on the display, asking you to remove all cables
and probes from oscilloscope.
49