HDS2062M-N Handheld DSO & DMM
7-Advanced Function of Oscilloscope
Figure
53: only under vertical mode multi-period waveform
Figure
54: only under horizontal mode multi-period waveform
Note:
1.
At the mode of Autoscale, the oscilloscope can self-estimate “Trigger mode”
(Edge, Video, and Alternate) and “Type” (Edge, Video). If now, you press
“Trigger mode” or “Type”, the forbidden information will display on the screen.
2.
At the mode of XY and STOP status, pressing
AUTO SET
to enter into
Autoscale, DSO switches to YT mode and AUTO status.
3.
At the mode of Autoscale, DSO is always in the state of DC coupling and AUTO
triggering. In this case, the forbidden information will be showing when making
Triggering or Coupling settings.
4.
At the mode of Autoscale, if adjust the vertical position, voltage division, trigger
level or time scale of CH1 or CH2, the oscilloscope will turn off Autoscale
function and if press AUTOSET again, the oscilloscope will enter into
Autoscale.
5.
Turn off the submenu at the Autoscale menu, the Autoscale is off and turn on the
submenu still enters into the function.
6.
When video triggering, the horizontal time scale is 50us. If one channel is
showing edge signal, the other channel is showing video one, the time scale
56